14
Table 4-A. System Difference to YSI within CGM
Glucose Ranges (Pediatric Study)
CGM
Glucose
Number
Ranges
of Paired
mg/dL
CGM-YSI
(mmol/L)
Overall
2922
*40-60
19
(2.2-3.3)
*61-80
76
(3.4-4.4)
81-180
1155
(4.5-10.0)
181-300
1380
(10.1-16.7)
301-350
206
(16.7-19.4)
351-400
86
(19.4-22.2)
* F or CGM ≤ 80 mg/dL (4.4 mmol/L), the differences in mg/dL (mmol/L)
are included instead of percent differences (%).
NOTE: CGM readings are within 40 to 400 mg/dL (2.2- 22.2 mmol/L),
inclusive.
164
|
Dexcom G4 PLATINUM (Pediatric) Continuous Glucose Monitoring System
Mean
Median
Percent
Percent
Difference
Difference
(%)
(%)
13.5
11.6
-18.1
-9.1
(-1.0)
(-0.5)
-3.7
-2.3
(-0.2)
(-0.1)
11.9
9.7
14.8
12.4
19.2
15.9
18.5
15.5
Mean
Median
Absolute
Absolute
Percent
Percent
Difference
Difference
(%)
(%)
17.4
13.5
19.2
9.1
(1.1)
(0.5)
13.4
10.6
(0.7)
(0.6)
17.0
13.0
17.4
13.3
19.4
15.9
19.1
15.5