Verification Tests for US (lb) Scales (Continued)
Table 4-4
Verification Test for US
Scale Applied Load lb
30.00
Over Capacity Test
Over Capacity 30.20
Decreasing Load Test
30.00
20.00
10.00
5.00
0.10
Return to Zero Test
0.00
Shift Test (see
Figure
4-4)
10.00
Position 1
10.00
Position 2
10.00
Position 3
10.00
Position 4
Verification Tests for Metric Scale
Table 4-5
Verification Test for
Metric Scale Applied
Load kg
Increasing Load Test
0.000
0.100
2.500
5.000
10.000
15.000
SCALE CALIBRATION (MODELS WITH A SCALE ONLY)
Indication = Applied
Load Within
Applicable
Tolerances
30.00
EEEE
30.00
20.00
10.00
5.00
0.10
0.00
10.00
10.00
10.00
10.00
Indication = Applied
Load Within
Applicable
Tolerances
0.000
0.100
2.500
5.000
10.000
15.000
All Tolerances
All Tolerances
Are +/-
Are +/-
Acceptance
Acceptance
Single Int
Dual Int
.015 lb
.015 lb
N/A
N/A
.015 lb
.015 lb
.01 lb
.01 lb
.01 lb
.005 lb
.005 lb
.005 lb
.005 lb
.0025 lb
0.0 lb
0.0 lb
.01 lb
.005 lb
.01 lb
.005 lb
.01 lb
.005 lb
.01 lb
.005 lb
All Tolerances
All Tolerances
Are +/-
Are +/-
Acceptance
Acceptance
Single Int
Dual Int
0.0 kg
0.0 kg
.0025 kg
.001 kg
.0025 kg
.002 kg
.005 kg
.003 kg
.005 kg
.005 kg
.0075 kg
.0075 kg
4 - 9