5. Installing the Scale

This procedure includes all of the steps for installing the HV-WP series and HW-WP
series. Therefore, on some products, there are some unnecessary steps.
Step 1 Connect the indicator unit to the column with
the accesory knobs and rubber washers.
Step 2 Take the base unit and column out, taking
care that the load-cell cable is not pulled.
Step 3 Put the weighing pan on the base unit.
Step 4 Insert the remainder of the load cell cable
into the column. Attach the column to the
bracket of the base unit so as not to damage
the load cell cable. Affix the column to the
bracket using two 6mm Allen screws.
Step 5 Select the place for installing the scale. Also
consider "4. Caution" on page 7.
Step 6 Adjust the level of the base unit by using the
"Bubble spirit level" and "Leveling foot".
Step 7 Ground the scale using the earth terminal.
Caution
Please confirm that the local voltage and
the receptacle type are correct for your
scale.
Step 8 Adjust the angle of the indicator unit using
the knobs on the side of indicator unit.
Step 9 Check the weighing accuracy. If the scale
needs calibration, refer to "14. Calibration".
on page 40.
Page 8
Column
Knob
Rubber washer
Pan
Step 3
Load-cell cable
Base unit
Bubble spirit level
Leveling foot
Step 5
OK
NG
Power label
Step6
Earth terminal
Power terminals
Step7
HVG-WP Series
Cable
Indicator
Column
Step 2
Earth
Earth
Knob