8. Counting Mode
The counting mode is the function to convert the total mass value (total weight) of
articles to a count by calculating a unit weight for each individual item.
It is necessary to store this unit weight to count articles.
8.1. Storing a Unit Mass
Step 1 Press the
Step 2 Press the
mass.
Step 3 Press the
The greater the quantity of samples, the greater the
accuracy of the count.
5 pieces, 10 pieces, 20 pieces, 50 pieces, 100 pieces
Step 4 Put a container item on the weighing pan.
Press the
Step 5 Put in the number of samples as selected at step 3.
Press the
stability mark is displayed. Then the count is
displayed.
Caution
When the sample is too light and it is not possible
to calculate a unit mass, the scale displays
and returns to step 3. The sample quantity needs
to be increased.
Pressing the
It will display the larger sample quantities.
Select a larger sample quantity and move to step
4.
key to display the unit
MODE
key to enter the mode to store a unit
SET
key to select the number of samples.
∧
key.
TARE
key to store the value after the
ENTER
key after
ENTER
.
lo ut
. is displayed.
lo ut
Page 22
HVG-WP Series