2.2. Installing the Balance

Install the balance as follows:
1. Refer to
"3.
Precautions" for installing the balance.
2. Refer to
"2.1.
Unpacking", to attach the pan support and the weighing pan on the weighing unit.
3. Insert the RS-232C cable into the jack located on the rear of the weighing unit.
When confirming the performance
of the weighing unit using WinCT-
AD4212C, connect the cable into
RS-232C terminal of the computer.
D-sub 9-pin
How to disconnect the cable from the weighing unit
Slide the connector sleeve in the direction of the arrow to unlock and gently pull the connector out.
Slide the connector sleeve
in the direction of the arrow
4. Confirm that the AC adapter type is correct for the local voltage and power receptacle type.
5. Plug the AC adapter plug into the AC adapter jack located on the rear of the weighing unit and plug
the AC adapter into the electrical outlet. Warm up the balance for 30 minutes or more with nothing
on the weighing pan.
6. Set the pan unit and I/O unit to adapt to the peripheral system. Set the following for the I/O unit.
 RS-232C (Refer
Set the weighing speed to adapt to the ambient conditions.
Note
Make sure that the personal computer is in operation before connecting to the AD-4212C.
7. After the balance has been installed, adjust the sensitivity of the balance using the 200 g calibration
weight provided with the balance. For details, refer to
(The AD-4212C-3000/6000/3100 can be adjust the sensitivity of the balance using the 200 g weight.)
8. By removing the provided weighing pan, a locally fabricated weighing pan can be attached to the AD-4212C
series by using the four screw holes (M4 screw x pitch 0.7 mm) on the pan support.
to"6. Communication
Rear of the
weighing unit
RS-232C cable
Unlocked
Specifications")
"4.2. Sensitivity
10
Arrow mark
face up
Adjustment".