Bright-Field Microscopy - Nikon ECLIPSE 80i Instructions Manual

Microscope
Hide thumbs Also See for ECLIPSE 80i:
Table of Contents

Advertisement

Chapter 3 Part Names and Microscopy Procedures(With DIH-M Mounted)

Bright-Field Microscopy

3.3
1
Turn on power.
The C-Box and the power source for the
Epi-illumination do not need to be turned on. If
they are turned on, keep the shutter of the
Epi-illumination closed.
(The shutter is automatically kept closed when the
power switch is turned on. Do not touch the hand
switch. If the shutter is opened accidentally, press
the hand switch once again to close the shutter.)
2
Press the preset switch.
3
Move the ND8, ND32, and NCB11
filters into the optical path.
4
Pull out both optical path
switching levers and direct the
full optical path toward the
binocular section.
3.3 Bright-Field Microscopy
The full optical
path directed
toward the
binocular
section
41
Press the power switch to
the "|" position.
Press the preset switch.
NCB11
ND32
ND8
Pull out both upper and
lower optical path
switching levers.
3

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents