Notes on handling the product

1. Handle the product gently.
This product is a precision optical instrument and requires gentle handling. Avoid subjecting it to sudden
impact and shocks.
Even relatively minor impacts are capable of affecting the precision of the object.
2. Weak electromagnetic waves
The product emits weak electromagnetic waves. Do not place precision electronic devices near the
product to avoid degrading their performance. If TV or radio reception is affected, move the TV or radio
farther away from the product.
3. Scratches, dirt, and foreign particles on the lens
Scratches or fouling such as fingerprints on optical components (such as lens and filters) will degrade
microscope images. If these parts become dirty, clean them as described in chapter "7. Care and
maintenance" at the end of this manual.
4. Dirt on the lamps
Never touch the lamp with bare hands. Dirt or fingerprints on the lamp will result in uneven illumination
and reduce the service life of the lamp. Always wear gloves when handling lamps.
5. Installation location
This product is a precision instrument. Use or storage in inappropriate environments may result in
malfunctions or poor performance. Consider the following factors when selecting an installation location:
• Select a vibration-free location. Install the product on a level surface.
• Install the product at least 10 cm away from walls.
• Choose a location less exposed to hazards in the event of collisions, earthquakes, or other potential
disasters. To keep the product from falling, use strong rope or other means if necessary to secure it
to the working desk or to another heavy, stable item. To fix the microscope, there are two M6 screw
holes (one at the right and the other at the left) behind the recess at the base of the main body
(back side).
• Avoid locations exposed to direct sunlight, locations immediately under room lights, and other bright
locations.
• Avoid locations with excessive dust.
• Do not install the product near any liquid.
• Make sure the ambient temperature is 0 to 40°C and humidity is 85% or less. Installing the 80i in hot,
humid locations may result in mildew formation or condensation, impairing performance or
generating malfunctions.
• Do not install the product in a locker or cabinet.
• Select a layout that allows easy removal of the power cord from the product's AC inlet in the event of
an emergency.
• Room lights just above the microscope may reduce visibility in the objective as extraneous light. If
possible, switch off room lights directly above the microscope when making observations.
Notes on handling the product
7