CHAPTER 3 MAINTENANCE PROCEDURES
3.3.18 Reassembling Analytical Load Cells
17. Install the top bracket. (See Figures 3-92 and 3-93.)
Figure 3-92. Analytical InCal Load Cell
with Top Bracket visible.
18. Remove the screws and shims
from the Hanger.
19. Install the Bottom Mounting Plate
as shown. Note the position of the
Weigh Below Hook. (See Figure
3-94.)
20. Position the Load Cell in the
balance Housing. Insert and
tighten the three screws, from
underneath the Bottom Housing.
Be sure the Load Cell is squarely
mounted in the Housing. Connect
the cable from the Main PCB to the
Position Sensor Board. (See
Figure 3-95.)
Ohaus Corporation www.ohaus.com
Top Bracket (with
InCal Weight Arms)
Screws on
Bottom Mounting
Plate
Cable from Main
PCB to Position
Sensor Board
3-32
Figure 3-93. Analytical balance Load Cell
with Top Bracket visible.
Figure 3-94. Bottom Mounting Plate on
Analytical Load Cell.
Figure 3-95. Ribbon cable connects PSB to
Main PCB.
Pioneer Balance Service Manual
Top Bracket