3.3.18 Reassembling Analytical Load Cells

The Position Sensor Board contains an Optical Sensor which must be positioned in
the Ratio Beam's center of travel for the balance to function properly.
21. Adjust the Position Sensor Board: power on the balance with the Top Housing off but
the Weighing Pan in its normal position. The On/Off switch is the second button
from the right on the main PCB.
22. Adjust the vertical position of
the Position Sensor Board: if it
is too high or too low, the
display indicates Error 8.X.
Move it until the display
indicates other than 8.X.
Then tighten the screws on
the board. (See Figure 3-96.)
23. Install the Load Cell Cover.
24. Perform Off-Center Load Adjustments. (See Chapter 4.)
25. Test the balance. (See Chapter 4.)
Pioneer™ Balance Service Manual
CHAPTER 3 MAINTENANCE PROCEDURES
NOTE:
Figure 3-96. When Position Sensor Board shows
normal weight, tighten its screws with an Allen Key.
33
3-
Ohaus Corporation www.ohaus.com