Off-Center Load Test - OHAUS PA64 Service Manual

Hide thumbs Also See for PA64:
Table of Contents

Advertisement

4.4.1 Precision Test – Precision and Analytical Balances
5. If the deviation for any set of readings (using the same mass placed on the center of
the Pan) is greater than the tolerance listed in Table 4-1 or 4-2, the balance does not
meet the precision specification. Inspect and correct the following areas:
5.1. Check for mechanical obstructions. Any foreign object touching any part of the
moving assemblies will cause a balance to fail the Precision Test. Inspect and
correct as necessary.
5.2. An error in the Off-Center Load Adjustment can affect the results of the
Precision Test. Inspect and correct if necessary. See the next section.

4.4.2 Off-Center Load Test

The Off-Center Load Test is used to determine whether displayed weight values are affected by
moving the sample to different areas of the Pan. See Figure 4-2.
1. Place a mass weighing half of the balance's capacity in the center of the Pan.
2. Press the On/Off button to return the reading to zero.
3. Move the mass halfway to the rear of the Pan and note the reading.
4. Move the mass halfway between the center and the left edge of the Pan and note the
reading.
5. Repeat this test for the right and front edges and note the readings.
Note any differences in the displayed weight reading at all positions. See Tables 4-2
and 4-3 for maximum allowable variance for each of the four positions.
Pioneer™ Balances Service Manual
Mass Locations
Figure 4-2. Mass Placement Locations
for Off-Center Load Test.
4-3
CHAPTER 4 TESTING
Ohaus Corporation www.ohaus.com

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents