2.2.2 Checking Load Cells for Trouble
Measure the voltages on +SIG and –SIG wires, disconnected from PCB.
Note: Measurements must be made with these wires disconnected from the
PCB. These measurements represent the output of the Load Cell. Record
measurements at Zero Load, 50% and full scale capacities. See Table 2-3 for
typical readings.
NOTE: Table 2-3 indicates typical readings. Actual values can vary, but should
remain linear throughout the range. If readings are out of tolerance, replace the
Load cell. (See Section 3.6.)
TABLE 2-3. LOAD CELL OUTPUT READINGS (in mV with 5V Excitation)
Model/Capacity
V21P3 / 3kg
V21P6 / 6kg
V21P15 / 15kg
Valor™ 2000 Series Service Manual
AC &
Battery
Connector
1 2 3 4
5 6 7
TOP VIEW OF PCB
1 2 3 4
Load
Cell
Figure 2-1. Valor 2000 PCB connections
Zero Load
0.8 ± 2.0
0.9 ± 2.0
0.4 ± 2.0
2-3
CHAPTER 2 DIAGNOSTIC GUIDE
Rear Display
PCB Connector
(Not used)
1 2 3 4 5
1 2 3 4 5
50% Load
Zero + 3.8
Zero + 4.7
Zero ± 4.1
Ohaus Corporation www.ohaus.com
100% Load
Zero + 6.8
Zero + 7.6
Zero ± 7.8