CHAPTER 1 GETTING STARTED
1.7

RS232 Commands

Commands listed in the following table will be acknowledged by the balance. The balance will
return "ES" for invalid commands.
Command
Function
IP
Immediate print the displayed weight (stable or unstable).
P
Print and displayed weight (stable or unstable).
CP
Continuous Print. OP ends continuous print.
AS
Auto print on stability
xP
Interval Print x = Print Interval (1-3600 seconds) OP ends Interval Print.
T
Same as pressing Zero/Tare key.
PU
Print current weighing unit.
OFF
Turns balance OFF.
ON
Turns balance on.
PSN
Print Serial Number.
PV
Print version: name, software revision and LFT ON (if LFT is set on).
x#
Set PC reference weight (x) in grams. (Must have been a APW stored.)
P#
Print PC reference weight.
x%
Sent % reference weight (x) in grams. (Must-have reference weight stored.)
P%
Print percent reference weight.
0PF
Stop (zero) the old Adventurer print format, begin to use the default print format
1PF
Stop using the default print format, begin to use the old Adventurer print format
0S
(zero) Stable off
SS
Stable on (default)
Stops the scale from sending "OK" in response to instruction (non-print) commands.
0RL
Returns scale to default mode of sending "OK".
1RL
Ohaus Corporation www.ohaus.com
TABLE 1-10. RS232 COMMAND TABLE
1-10