3.3.8 Load Cell Disassembly – Precision Balances
15. Remove the Service Fixture,
releasing the Hanger.
(See Figure 3-28.)
16. Remove the two screws from the Magnet Half. (See Figure 3-29.) Pull the Magnet
Half away from the Ratio Coil. (Pull hard to overcome magnetic force and any
possible corrosion.)
17. Remove the screws in the Sideways Motion Stop Plate,
and lift it off. (See Figure 3-30.)
Figure 3-29. Two screws secure Magnet Half.
Pioneer™ Balance Service Manual
CHAPTER 3 MAINTENANCE PROCEDURES
Figure 3-28. Hanger.
Figure 3-30. Sideways Motion
3-11
Ohaus Corporation www.ohaus.com
Hole in vane must
be clear.
Stop Plate.