EN-1 Pioneer ™ 1. INTRODUCTION 1.1 Safety Precautions Please follow these safety precautions: • Verify that the AC Adapter input voltage matches the local AC power supply. • Use the balance only in dry locations. • Do not operate the balance in hostile environments.
Pioneer EN-2 ™ 2.2 Install Components 0.1g and 0.01g Models 1) Install Wind-Ring (InCal 2) Install Pan Support 3) Install Pan only) 0.001g and 0.0001g Models 1) Install Side Doors - Insert fully into Top Frame then down over retainer.
EN-3 Pioneer ™ 2.3 Level Balance Level the balance on a firm, steady surface. Avoid locations with excessive air current, vibrations, heat sources or rapid temperature changes. 2.4 Connect Power For use with CSA Certified (or equivalent approved) power source, which must have a limited circuit output.
Pioneer EN-4 ™ accuracy of the count. The balance automatically recalculates the Average Piece Weight when the number of parts on the pan is less than three times the original sample size. The display shows [APwW.OPT] each time the APW is optimized.
EN-5 Pioneer ™ .CAL. 4.2 Calibration Menu InCal or Span calibration should be performed daily and when the room temperature changes. InCal [InCal] calibrates the balance using an internal mass. Span Calibration [SpaN] uses two weight values: zero and a weight between 50% and 100% of the capacity of the balance.
Pioneer EN-6 ™ LSD – The Least Significant Digit (LSD) is the number of displayed divisions (d) by which the weight is incremented. The values are 1d, 2d, 5d, 10d, 100d or 0.5d. Press No to go to the next setting, press Yes to accept.
5.2 Service Information If the troubleshooting section does not resolve or describe your problem, contact your authorized Ohaus service agent. Please visit our web site, www.ohaus.com to locate the Ohaus office nearest you. 5.3 Accessories...
Pioneer EN-8 ™ 6. TECHNICAL DATA Ambient conditions – The technical data is valid under the following ambient conditions: Ambient temperature: 10°C to 30°C Relative humidity: 15 % to 80 % at 31°C non-condensing, decreasing linearly to 50% at 40°C Height above sea level: Up to 2000 m Operability is assured at ambient temperatures between 5°C and 40°C...
Page 10
AS/NZS4251.1 Emission; AS/NZS4252.1 Immunity CAN/CSA-C22.2 No. 1010.1-92; UL Std. No. 3101-1 ISO 9001 Registration In 1994, Ohaus Corporation, USA, was awarded a certificate of registration to ISO 9001 by Bureau Veritus Quality International (BVQI), confirming that the Ohaus quality management system is compliant with the ISO 9001 standard’s requirements.
Ohaus. In lieu of a properly returned warranty registration card, the warranty period shall begin on the date of shipment to the authorized dealer. No other express or implied warranty is given by Ohaus Corporation. Ohaus Corporation shall not be liable for any consequential damages.