7.4
Basic setup
In this section the settings required for using the device as prescribed are summarized. For more exten-
sive settings refer to the appendix.
7.4.1
Setting the overtemperature switch-off point T
Warning !
Hold the key
pressed during the complete setting procedure:
Tmax
Max. temperature
Max: 205
Min: 0
סּ
- - -
OK
On releasing
you are returned to the menu level without any change.
For T
the following applies: 5 K above desired bath temperature, but at least 25 K below the fire point
max
of the heat transfer liquid used.
46
Overheating due to entering an incorrect T
point temperature
•
Set T
in each case according to the heat transfer
max
liquid used (at least 25 K below fire point).
−
Press the enter key
The entry window appears. The cursor flashes under the T
value. The maximum and minimum adjustable temperature
values are displayed.
−
Change the value with
Note: With a longer depression the figures increment faster.
−
85
Single figures can be selected by pressing
−
Confirm your choice with the enter key
- - -
Operation
max
Burns, scalds, fire
.
ECO GOLD
and set-
max
or .
or .
.
25/08/2011/ YACE0088