Lauda E 4 G Operating Instructions Manual page 52

Heating and cooling thermostats with control head gold
Hide thumbs Also See for E 4 G:
Table of Contents

Advertisement

No.
Alarm
3
Overtemperature
3
Overtemperature
‫סּ‬
Display
OK
52
When the bath temperature is located above the overtemperature
switch-off point, a two-tone alarm sounds. "Overtemperature" ap-
pears in the display, the heater switches off on all poles and the
pump and chiller are switched off via the electronics.
Rectify the cause of the malfunction.
Wait until the bath temperature has cooled below the overtem-
perature switch-off point or set the overtemperature switch-off
point higher than the bath temperature.
If "Overtemperature" appears in the display:
Unlock the "Overtemperature" display with
Standby
Maintenance
ECO GOLD
.
25/08/2011/ YACE0088

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents