Setting The Temperature Target Value T; Basic Settings; Safety Mode - Lauda P 10 Operation Manual

Pro bath thermostats and circulation thermostats with base remote control
Hide thumbs Also See for P 10:
Table of Contents

Advertisement

5.10

Setting the temperature target value T

Fig. 34: Entering set temperature
5.11

Basic settings

5.11.1

Safety Mode

Purpose of Safety Mode
V04REV14
set
The temperature target value T
perature equipment should reach and then maintain.
This section is relevant for:
Device operation using the Base remote control unit
1.
Press any button on the Base unit to display the home window.
2.
Press the [input button] to open the menu.
3.
Press the [input button] again to select the first [Set temperature]
menu item.
An input window appears on the display.
4.
Adjust the new set temperature using the cursor buttons.
5.
Press [OK] to confirm the new value.
The value is accepted.
6.
Press the [ESC] softkey to open the home window.
Safety Mode gives users of LAUDA thermostatic circulators the option of
defining a safe temperature state in advance. In the event of a malfunction or
error, the user can manually switch to this state quickly and easily or allow the
device to do so automatically.
When Safety Mode is active, the thermostatic circulator continues to
operate at the set temperature T
the Safety Mode menu.
In the Safety Mode submenu, the user specifies how the device should
respond (actions) if certain events occur.
Which events lead to the activation of Safety Mode?
Being activated manually via the thermostatic circulator.
Command via interface.
Being triggered by certain alarms.
Communication with the control station being interrupted.
What indicates that Safety Mode is active?
A flashing Safety Mode icon
is operating. The pump impeller and pump level number are hidden.
If the device is on standby, T
displayed instead.
What happens in active Safety Mode if there is no alarm pending?
The Safety Mode window appears on the Base display and gives you the
option of quitting Safety Mode .
PRO bath thermostats and circulation thermostats
is the temperature that the constant tem-
set
and control variable previously defined in
set
appears on the display when the device
is hidden and the Safety Mode icon
ext
is