Entering the offset value
Activating an offset source
6.7

Limiting heating and cooling

92 / 171
1.
Select the [Diff. set/actual] button in the setpoint offset menu.
The entry window appears. An offset value can be entered within
the limits displayed.
2.
Enter the setpoint offset.
3.
Press the [Enter key] to confirm.
4.
The software returns to the previous Setpoint offset menu.
You can activate or deactivate the value entered for the setpoint offset
of a corresponding source using the options in the [Offset source] menu.
[External Pt100], for example, allows you to activate the setpoint offset for
the external temperature probe.
1.
Select the [Offset source] button in the Setpoint offset menu.
2.
Select one of the following options:
Select [Off] to deactivate the offset source.
n
Activate an offset source from the remaining options:
[Extern Pt100]
n
[Extern analog]
n
[Extern serial]
n
[Extern USB]
n
[Extern Ethernet]
n
[Extern EtherCAT]
n
[Extern Pt100-2]
n
3.
Press the [Enter key] to confirm.
4.
Press the [ESC] softkey to switch to the home window.
The maximum heating power or max. cooling capacity can be limited using
the controller output limit. The limit is set as a percentage of the maximum
value.
The controller output limit for the heating output is designed to prevent
excessive temperatures on the surface of the heater. Excessive heater tem-
peratures may degrade the heat transfer liquid.
1.
Press the [Enter key] to open the menu.
2.
Select the menu items  Setup  Control  Cont.outp.limit.
3.
Select [Max. heating] and press [OK] to confirm.
An entry window appears. The controller output limit can be set to
a value within the limits displayed.
4.
Change the value accordingly.
5.
Press the [OK] button to return to the previous screen with the new
setting configured.
The new setting is active.