ECO GOLD
c) Cooling bath and circulation thermostats
The device can be operated up to an ambient temperature of 40 °C.
An increased ambient temperature results in reduced cooling power.
When putting the chiller into operation after a lengthy shut-down period, up to 30 minutes
may pass until the rated cooling power is available depending on room temperature and
device type.
28.04.10/ YACE0088
Do not tilt the device and never turn it upside
down.
If possible after transport, site the device two
hours before putting it into operation.
Do not cover the ventilation openings at the
back of the device and on the lower part of the
device.
Keep a distance of at least 40 cm free on all
sides of the device.
Set the flow distribution to INT (
during operation as a bath thermostat (without
external load) the flow is discharged from the
opening for the internal bath circulation.
During operation as a bath thermostat without
an external load and with the pump connec-
tion set fitted, the outflow nozzle of the pump
connection set must be closed (use sealing
plug) or short circuited with the return nozzle.
For bath temperatures above 70 °C attach the
sticker
items to an easily visible point on the bath.
Operation with external load (circulation ther-
mostat) (
Preparations
included in the supplied
6.2).