Lauda E 4 G Operating Instructions Manual page 61

Heating and cooling thermostats with control head
Hide thumbs Also See for E 4 G:
Table of Contents

Advertisement

ECO GOLD
Extern Pt100
Kpe
Tne
Xpf
Tv manual/auto
Tve
Tde
Prop_E
‫ סּ‬OK
ESC
Kpe
Max: 50
Min: 0,1
‫ סּ‬OK
ESC
28.04.10/ YACE0088
Access to the menu level is obtained by pressing
0.5
Selection and confirmation of
100
Control
4.0
cent menu window appears. Apart from the control parame-
auto
ters the currently set values are displayed.
83(auto)
Under the menu point "Tv manual/auto" you can select
83(auto)
20(auto)
between manual and automatic entry using
lection "automatic" is displayed in the menu line by (auto).
Standby
If "automatic" is selected, the entry is blocked for the pa-
rameters Tv and Td.
Select the parameter with
The respective edit window appears with Min and Max
figures for the parameter values Kpe, Tne, Tve, Tde and
Xpf.
By pressing
ally and change them with
0,5
You confirm the set value by pressing
- - -
You quit the window without changes with
Putting the device into operation
Control parameter
.
or
you can also select numbers individu-
Setup
extern
Pt100. The adja-
. The se-
or
and confirm it with
or
.
.
(ESC).
.