Type the default procedure name to be used by the
system here. The default procedure name can be
composed of up to 10 alphanumerical characters.
Note
• To prevent the names of the image files on the
UPA-P100MD from being truncated when you view
them in the browser file list, use a short default
doctor name and procedure name.
• Also note that when you specify a default doctor
name and/or default procedure name, and select
[Use default], the default names appear on the RM-
P110 Remote Control Unit display, where you can
modify them if necessary. If you do not need to
make any modifications, simply press the [ENT]
key to proceed. See "Grabbing Images" for details
about entering names using the remote control.
You need to select the local time zone here. Select the
time zone in your area from the list.
Set the date and time in these boxes.
After setting the date and time, be sure to select this
box. Otherwise, the new settings are not applied
when you click [Save Config].
Select the language that you want to use for the
remote control display. You can choose one of 12
languages from the list as the display language. By
default, this language is English. Also, note that the
configuration screen language is not affected by this
setting.
Be sure to select this check box when you want to
capture images from an analog printer synchronized
with the UPA-P100MD. The capture of images is
delayed by the time specified in the [Delay value (in
milliseconds)] box. By default, this is 250.
In this box, enter the delay capture time in
milliseconds. By default, this is 250.
Enter the minimum length of time during which the
switch of an external control unit connected to the
UPA-P100MD parallel port must be depressed. The
shorter the time set here, the faster the key response.
By default, this is 30.
Click this button to save the changes that you make in
the configuration screen. The device restarts
automatically when you click [Save Config].
Note
To access the configuration screen once more, you
need to type the address specified in the [IP Address]
box (i.e. http://network IP address/) in the [Address]
box of you Internet browser, click [Configuration],
then click [Refresh]. Do not click [Back] on your
browser.
Click this button to cancel all the changes that you've
made in the configuration screen. The configuration
screen settings return to those at the time of opening
the configuration screen.