Oven Parameters
OPERATING SEQUENCE

Setting Up a Single Ramp Temperature Program

This program raises the initial oven temperature to a specified final temperature at
a specified rate and maintains the final temperature for a specified time.
Use INC and DEC keys to increase or decrease the parameter values and confirm
pressing ENTER.
1. In Main Menu move the cursor until
2. Scroll to
3. Scroll to
4. Scroll to
Instruction Manual
°C
300
200
Initial
time
(14 min)
100
Temp
(40 °C)
0
0
0
0
Figure 4-2. Example of Single Ramp Temperature Program
the
menu.
OVEN
and enter the initial the initial temperature.
Temp
Initial Time
the initial temperature.
and set it
Ramp 1
reach the ramp's Final Temp.
Ramp 1
(10 °C/min)
10
10
20
20
30
30
is selected. Press ENTER to display
Oven
and enter the time you want the oven to maintain
. Enter the ramp rate in °C/min for the oven to
On
Chapter 4
Column Oven
Final
temp 1
(300 °C)
Final
time 1
(10 min)
40
40
min.
min.