Carrier Gas Parameters
This paragraph explains the electronic programming and control of the GC carrier
gases. The Digital Pressure Flow Control (DPFC) module regulates carrier gas
electronically. You control the gas flow by programming parameters in the carrier
gas menu.
The
of the carrier gas type. Parameters change according to the selected flow mode:
constant flow or constant pressure.
In Main Menu, move the cursor until
the
Flow Mode
Two flow modes are available.
Constant Flow
In constant flow mode, the column flow is kept constant throughout the analysis.
The pressure at the column head will change with the column temperature to
maintain the set flow.
Constant Pressure
In constant pressure mode, the pressure at the column head is kept constant
throughout the analysis. Flow rate will decrease as oven temperature increases.
80
menu includes the control parameters for the carrier gas, regardless
CARRIER
menu.
CARRIER
Carrier
CARRIER
Flow mode
Const flow