Adjusting the fitting controls
By clicking the <> each or all of the three regions can be
expanded to display all the underlying frequency channels.
Click to display the MPO controls for each region or channel.
Adjust the gain controls in 1 dB steps in the selected fields.
MPO is displayed in dBHL while gain is displayed in dB (IG).
14 VIGO CONNECT FIT TING GUIDE
RITE/Corda
2
only:
For RITE instruments and BTE 312 or 13 instruments with Corda
with domes no low-frequency gain is provided in the initial
setting due to the dominating natural sound.
Therefore the LF region is disabled in the fitting. If you want to
make adjustments in the LF region, expand the region and
you will get access to the underlying frequency channels.

Program Manager

2
In the Program Manager you can activate, configure, delete,
and change the order of the programs.
For general programs you can choose between the NAL NL1
and DSL v5.0 fitting rationales.
Two phone programs exist, Microphone (M) or telecoil (T).
You choose whether you want to activate the program on the
left or right ear. By default a copy of P1 reduced will be placed
in the other instrument.
If you have an Auto Phone in the instrument, you can activate
the automatic functionality in the lower right corner.
You do not have to add a Phone program in P2 – P4. You can
choose whether the program should be a Microphone or a
Telecoil program (when Telecoil is available). This program is
configured either here or in the End Fitting Step.
If you are fitting Streamer to your Vigo Connect instruments,
adjustments to Streamer programs are done in the End Fit-
ting step.
BTE instruments are compatible with DAI and FM shoes and
RITE instruments are compatible with the DAI shoe. When the
shoe is attached to the instrument, you will automatically
have access to two extra programs:
• DAI/ FM + Microphone (default) and
• DAI/FM only.
The two programs are based on the settings in P1.
Any adjustments are handled in the End Fitting step.