Operation
3.5.2

Voltage Spike Example

This example illustrates programming the SW for a voltage spike on a 150V sine
waveform that starts at 30° and ends at 50° with an amplitude of 220.6 volts.
MAIN MENU
PROGRAM
MEASURE
WAVEFORM
SEQUENCE
INSTR
SYSTEM
1.
At the MAIN MENU, select the WAVEFORM menu, then press <Enter>.
WAVEFORM
LOAD
EDIT
VIEW
SAVE
DELETE
2.
The WAVEFORM Menu will default to the LOAD sub-menu. If the cursor is not
pointing to LOAD, use the Knob or Up/Down Arrow keys to select LOAD.
3.
Press the Enter key to enter the FUNC (function) mode.
4.
Press <Enter> again to place a box around the waveshape selection.
5.
Use the Knob to cycle through the waveshape selections until SINE appears in
the FUNC box.
3-38
SW 5250A•SW 3500A•SW 1750A
OUTPUT [OFF]
OUTPUT [OFF]
Operation Manual