Elgar SmartWave SW5250A Operation Manual page 55

Switching amplifier
Table of Contents

Advertisement

SW 5250A•SW 3500A•SW 1750A
3.3.4.2
EDIT Sub-Menu
The WAVEFORM menu allows you to EDIT a waveform in the scratchpad.
WAVEFORM
LOAD
EDIT
VIEW
SAVE
DELETE
EDIT Sub-Menu Status and Information Data:
Menu Item
Status and Information Line
FREQ
[waveform frequency]
Vrms
[output waveform voltage]
START
[starting phase angle]
TIME
[surge/sag duration]
STOP
[ending phase angle]
AMPL
[drop/spike voltage]
NOTE: Vrms is in terms of RMS voltage while AMPL is in terms of peak voltage.
Both of these values depend on the selected voltage range setting in
the INSTRUMENT menu.
See the example in section 3.5.1.
Operation Manual
OUTPUT [OFF]
FREQ
Vrms
START
TIME
STOP
AMPL
Range
40–5000 Hz
0–156 or 0–312 V
0–360°
0–1/Freq
Start–360°
-220.6 to 220.6 Vpeak (156 Vrms range)
-441.2 to 441.2 Vpeak (312 Vrms range)
Operation
Hz
V
Deg
ms
Deg
V
3-11