Operation
3.3.7.5
CONFIG Sub-Menu
Pressing <Enter> to select CONFIG will display the System Configuration options.
BOOST - Allows the Servo Boost circuit to be disabled during non-sequence operation.
This option is important when running a sequence at low voltage levels. See Section
1.12.2, Servo Boost, for a full explanation of the Boost function. The factory default
state is Boost ON. This option should be set to OFF when:
The RMS Servo is ON (Low Frequency = OFF) and
Sequence mode is being used with Exit mode set to "Program Menu."
RESET CAL – Resets calibration constants to default values. Enter a confirmation code
(2096) to activate this option. If you enter an incorrect code, the unit will beep and the
message "! INCORRECT RESET CODE" will be displayed, the code will be reset to 0,
and no changes will occur to the calibration constants.
T&MB OPT – Displays firmware configuration for the Test & Measurement option. ON
indicates the Test & Measurement board is installed; OFF indicates it is not. This field is
for display purposes only.
SWAE OPT – Displays firmware configuration for the Low Impedance option. ON
indicates the Low Impedance hardware is installed; OFF indicates it is not. This field is
for display purposes only.
SYSTEM
USER
COMM
EXTERNAL
FAULTS
CONFIG
CONFIG Sub-Menu Status and Information Data:
Menu Item
BOOST
RESET CAL
T&MB OPT
SWAE OPT
3-32
BOOST
RESET CAL
T&MB OPT
SWAE OPT
Status and Information Line
[System boost enable/disable]
[Enter reset code]
[T&MB board installed?]
[Low impedance installed?]
SW 5250A•SW 3500A•SW 1750A
OUTPUT [OFF]
ON
0
OFF
OFF
Operation Manual