Table D-1 (cont.)
Models 196/199 device-dependent command summary
Mode
Filter
Rate
Trigger mode
Reading mode
Data store size
Interval
Value
V±nn.nnnn or
V±n.nnnnnnE+n
Default conditions
Data format
SRQ
Command
Description
P0
Filter disabled
P1
Moving filter (count = 10)
P2
Repeat filter (count = 10)
S0
0.1 PLC integration
S1
Line cycle integration (16.67ms, 60Hz; 20ms, 50Hz)
S2
10 PLC (166.67ms integration, 60Hz;
200ms integration, 50Hz)
T0
Continuous on Talk
T1
One-shot on Talk
T2
Continuous on GET
T3
One-shot on GET
T4
Continuous on X
T5
One-shot on X
T6
Continuous on External Trigger
T7
One-shot on External Trigger
B0
Readings from A/D converter
B1
Individual readings from data store
B2
All readings from data store (buffer dump)
I0
Disable data store
In
Data store of n (n=1 to 500), fill and stop
Q0
Default interval, 175ms (SELECT OFF)
Qn
n=interval in milliseconds (15ms to 999999ms)
Zero value, simulated reference junction temperature
L0
Restore factory default conditions
G0
Reading with prefix.
G1
Reading without prefix.
G2
Reading and buffer location with prefix.
G3
Reading and buffer location without prefix.
G4
Reading and channel with prefix.
G5
Reading and channel without prefix.
G6
Reading, buffer location, and channel with prefix.
G7
Reading, buffer location, and channel without prefix.
M0
Disable
M1
Reading overflow
M2
Data store full
M4
Data store half full
M8
Reading done
M16
Ready
M32
Error