Keithley 2010 User Manual page 189

Hide thumbs Also See for 2010:
Table of Contents

Advertisement

Description
This command is used to specify the simulated reference temperature. The
temperature value depends on which temperature scale is presently selected
(°C, °F, or K). Typically, 0° or 23°C is used as the simulated reference
temperature.
:REAL:TCOefficient <n>
[:SENSe[1]]:TEMPerature:TCouple:RJUNction[1]]:REAL:TCOefficient <n>
Parameters
<n> = -0.09999 to 0.09999
Query
:TCOefficient?
:TCOefficient? DEFault
:TCOefficient? MINimum Query lowest allowable TC
:TCOefficient? MAXimum Query largest allowable TC
Description
This command is used to specify the temperature coefficient (TC) of the
"real" temperature reference junction. TC is specified in °C/volt and is not
affected by the :UNIT:TEMPerature command.
:REAL:OFFSet <n>
[:SENSe[1]]:TEMPerature:TCouple:RJUNction[1]:REAL:OFFSet <n>
Parameters
<n> = -0.09999 to 0.09999
Query
:OFFSet?
:OFFSet? DEFault
:OFFSet? MINimum
:OFFSet? MAXimum Query largest allowable voltage offset
Description
This command is used to specify the offset voltage at 0°C for the specified
reference junction.
DEFault
MINimum
MAXimum
Query temperature coefficient (TC)
Query *RST default TC
DEFault
MINimum
MAXimum
Query voltage offset
Query *RST default voltage offset
Query lowest allowable voltage offset
SCPI Command Reference
Specify temperature coefficient
+0.00020 temperature coefficient
-0.09999 temperature coefficient
+0.09999 temperature coefficient
Specify voltage offset at 0°C
0.05463
-0.09999
0.09999