Keithley 2010 User Manual page 157

Hide thumbs Also See for 2010:
Table of Contents

Advertisement

Table 5-10
Trigger command summary
Command
:INITiate
[:IMMediate]
:CONTinuous <b>
:CONTinuous?
:ABORt
:TRIGger
[:SEQuence[1]]
:COUNt <n>
:COUNt?
:DELay <n>
:AUTO <b>
:AUTO?
:DELay?
:SOURce <name>
:SOURce?
:TIMer <n>
:TIMer?
:SIGNal
:SAMPle
:COUNt <NRf>
:COUNt?
Notes:
1. Defaults for continuous initiation:
:SYSTem:PRESet enables continuous initiation.
*RST disables continuous initiation.
2. Defaults for count:
:SYSTem:PRESet sets the count to INF (infinite).
*RST sets the count to 1.
Description
Subsystem command path:
Initiate one trigger cycle.
Enable or disable continuous initiation.
Query continuous initiation.
Reset trigger system.
Subsystem command path:
Path to program Trigger Layer:
Set measure count (1 to 9999, or INF).
Query measure count.
Set delay (0 to 999999.999 sec)
Enable or disable auto delay.
Query state of delay.
Query delay.
Select control source (IMMediate, TIMer,
MANual, BUS, or EXTernal).
Query control source.
Set timer interval (0.001 to 999999.999 sec).
Request the programmed timer interval.
Loop around control source.
Specify sample count (1 to 1024).
Query sample count.
SCPI Command Reference
Default
parameter
(Note 1)
(Note 2)
0
ON
IMMediate
0.1
1
5-19
SCPI