5-42
SCPI Command Reference
:AUTO <b>
[:SENSe[1]]:CURRent:AC:RANGe:AUTO <b>
[:SENSe[1]]:CURRent[:DC]:RANGe:AUTO <b>
[:SENSe[1]]:VOLTage:AC:RANGe:AUTO <b>
[:SENSe[1]]:VOLTage[:DC]:RANGe:AUTO <b>
[:SENSe[1]]:RESistance:RANGe:AUTO <b>
[:SENSe[1]]:FRESistance:RANGe:AUTO <b>
[:SENSe[1]]:VOLT[:DC]:STERminals:RANGe:AUTO <b>
Parameters
Query
Description
<b> =
1 or ON
0 or OFF
:AUTO?
These commands are used to control auto ranging. With auto ranging
enabled, the instrument automatically goes to the most sensitive range to
perform the measurement.
The auto range command (:RANGe:AUTO) is coupled to the command that
manually selects the measurement range (:RANGe <n>). When auto range is
enabled, the parameter value for :RANGe <n> changes to the automatically
selected range value. Thus, when auto range is disabled, the instrument
remains at the automatically selected range. When a valid :RANGe <n>
command is sent, auto ranging disables.
Control auto range for ACI
Control auto range for DCI
Control auto range for ACV
Control auto range for DCV
Control auto range for Ω2
Control auto range for Ω4
Control auto range for sense terminals
Enable auto range
Disable auto range
Query auto range (on or off)