Configuration - Keithley 2010 User Manual

Hide thumbs Also See for 2010:
Table of Contents

Advertisement

2010 Low-Noise Multimeter
FREQUENCY AND PERIOD CHARACTERISTICS
ACV
FREQUENCY
RANGE
RANGE
100 mV
3 Hz
to
to
750 V
500 kHz
FREQUENCY NOTES
1. Specifications are for square wave inputs only. Input signal must be >10% of ACV range. If input is
<20mV on the 100mV range then the frequency must be >10Hz.
2. 20% overrange on all ranges except 750V range.
TEMPERATURE CHARACTERISTICS
THERMOCOUPLE
2, 3, 4
TYPE
RANGE
RESOLUTION
J
–200 to + 760°C
K
–200 to +1372°C
N
–200 to +1300°C
T
–200 to + 400°C
4-WIRE RTD
2, 3, 7, 8
RANGE
RESOLUTION
–100° to +100°C
0.001°C
–200° to +630°C
0.001°C
Temperature Notes
1. For temperatures <–100°C, add ±0.1°C and >900°C add ±0.3°C.
2. Temperature can be displayed in °C, K or °F.
3. Accuracy based on ITS-90.
4. Exclusive of thermocouple error.
5. Specifications apply to channels 2–6. Add 0.06°C/channel from channel 6.
6. Excluding probe errors.
7. 100Ω platinum, D100, F100, PT385, PT-3916 or user type.
8. Maximum lead resistance (each lead) to achieve rated accuracy is 5Ω.
INTERNAL SCANNER SPEED
MAXIMUM INTERNAL SCANNER RATES: RANGE: CHANNELS/s
TRIGGER DELAY = 0
2 WIRE
DCV
2
ACV
2, 3
OHMS
All : 105
All : 96
All : 102
TRIGGER DELAY = AUTO
2 WIRE
DCV
2
ACV
2, 3
OHMS
100 Ω : 82
0.1 V : 100
All : 1.8
1 V : 100
1 kΩ : 85
10 V : 100
10 kΩ : 42
100 V : 70
100 kΩ : 28
1000 V : 70
1 MΩ : 8
10 MΩ : 5
100 MΩ : 3
Internal Scanner Speed Notes
1. Speeds are for 60Hz or 50Hz operation using factory default operating conditions (*RST). Auto Zero off,
Auto Range off, Display off, sample count = 1024.
2. NPLC = 0.01.
3. DETector BANDwidth: 300.
1,2
RESOLUTION
PERIOD
GATE
±(ppm of
RANGE
TIME
reading)
333 ms
to
1 s
0.3
2 µs
ACCURACY
90 DAY/1 YEAR (23°C ± 5°C)
RELATIVE TO SIMULATED
REFERENCE JUNCTION
0.001°C
±0.5°C
0.001°C
±0.5°C
0.001°C
±0.5°C
0.001°C
±0.5°C
90 DAY/1 YEAR
(23°C ± 5°C)
6
ACCURACY
±0.08°C
±0.14°C
1
4 WIRE
T/C
2
OHMS
2
TEMPERATURE
<10MΩ : 55
All : 70
4 WIRE
T/C
2
OHMS
2
TEMPERATURE
100 Ω : 42
All : 70
1 kΩ : 42
10 kΩ : 25
100 kΩ : 21
1 MΩ : 7
10 MΩ : 5
100 MΩ : 3
Triggering and Memory
ACCURACY
READING HOLD SENSITIVITY: 0.01%, 0.1%, 1%, or 10% of reading.
90 Day/1 Year
±(% of reading)
TRIGGER DELAY: 0 to 99 hrs (1ms step size).
EXTERNAL TRIGGER DELAY: <1ms.
0.01
EXTERNAL TRIGGER JITTER: <500µs.
MEMORY: 1024 readings.
Math Functions
Rel, Min/Max/Average/StdDev (of stored reading), dB, dBm, Limit Test,
%, and mX+b with user defined units displayed.
dBm REFERENCE RESISTANCES: 1 to 9999Ω in 1Ω increments.
1
REMOTE INTERFACE
USING
2001-TCSCAN
5
Keithley 199/196 Emulation
±0.65°C
GPIB (IEEE-488.2) and RS-232C
±0.70°C
SCPI (Standard Commands for Programmable Instruments)
±0.70°C
±0.68°C
2 YEAR
POWER SUPPLY: 100V / 120V / 220V / 240V.
(23°C ± 5°C)
6
ACCURACY
LINE FREQUENCY: 50Hz to 60Hz and 440Hz, automatically sensed at
power-up.
±0.12°C
±0.18°C
POWER CONSUMPTION: 22VA.
OPERATING ENVIRONMENT: Specified for 0°C to 50°C. Specified to
80% R.H. at 35°C. Altitude up to 2000 meters.
STORAGE ENVIRONMENT: –40°C to 70°C.
WARRANTY: 3 years.
SAFETY: Conforms to European Union Directive 73/23/EEC
EN61010-1, CAT II.
EMC: Complies with European Union Directive 89/336/EEC,
EN61326-1.
VIBRATION: MIL-PRF-28800F Class 3 Random.
WARMUP: 2 hours to rated accuracy.
DIMENSIONS:
Rack Mounting: 89mm high × 213mm wide × 370mm deep (3
8
3
Bench Configuration (with handle and feet): 104mm high ×
RTD
238mm wide × 370mm deep (4
2
TEMPERATURE
2
SHIPPING WEIGHT: 5kg (11 lbs).
All : 2
VOLT HERTZ PRODUCT: ≤8 × 10
ACCESSORIES SUPPLIED: Model 1751 Safety Test Leads, User
RTD
Manual, Service Manual.
2
TEMPERATURE
2
All : 2
A C C E S S O R I E S A V A I L A B L E
1050
1754
2000-SCAN
2001-TCSCAN 9-Channel Thermocouple Scanner
2010-EW
4288-1
4288-2
5804
5805
5806
5807-7
7007-1
7007-2
7009-5
8502
8503
8605
8606
GENERAL
in × 14
9
in).
8
1 6
in × 9
1
8
7
V·Hz.
Padded Carrying Case with handle and should strap
Universal Test Lead Kit
10-Channel Scanner
(includes 1-channel reference junction)
1-Year Warranty Extension
Single Fixed Rack Mount Kit
Dual Fixed Rack Mount Kit
4 Terminal Test Lead Set
Kelvin Probes
Kelvin Clip Lead Set
Helical Spring Point Test Leads
Shielded GPIB Cable, 1m (3.2 ft)
Shielded GPIB Cable, 2m (6.5 ft)
Shielded RS-232 Cable 1.5m (5 ft)
Trigger Link Adapter to 6 female BNC connector
Trigger Link Cable to 2 Male BNCs, 1m (3.2 ft.)
High Performance Modular Test Leads
High Performance Probe Tip Kit
in ×
1
2
in × 14
in).
3
9
8
1 6
HW 9/5/03
Rev. C

Advertisement

Table of Contents
loading

Table of Contents