5-50
SCPI Command Reference

Thermocouple commands

:TYPE <name>
[:SENSe[1]]:TEMPerature:TCouple:TYPE <name>
Parameters
Query
Description
:RSELect <name>
[:SENSe[1]]:TEMPerature:TCouple:RJUNction[1]:RSELect <name>
Parameters
Query
Description
:SIMulated <n>
[:SENSe[1]]:TEMPerature:TCouple:RJUNction[1]:SIMulated <n>
Parameters
Query
<name> =
J
K
N
T
:TYPE?
This command is used to configure the Model 2010 for the thermocouple
type that you are using to make temperature measurements.
These commands are used to configure the reference junction for
thermocouple temperature measurements.
<name> =
SIMulated
REAL
:RSELect?
This command is used to specify the type of reference junction that is going
to be used for thermocouple temperature measurements. Specify REAL if
you are using an actual reference junction. The :REAL command is then used
to specify the desired reference temperature. Specify SIMulated if you wish
to use a simulated reference temperature. The :SIMulated command is then
used to specify the desired simulated reference temperature.
<n> =
0 to 50
32 to 122
273 to 323
DEFault
MINimum
MAXimum 50°C, 122°F, 323K
:SIMulated?
:SIMulated? DEFault
:SIMulated? MINimum
:SIMulated? MAXimum
Specify TC type
Set operation for Type J thermocouples
Set operation for Type K thermocouples
Set operation for Type N thermocouples
Set operation for Type T thermocouples
Query thermocouple type
Specify reference junction type.
Use simulated temperature as reference
Use a measured temperature as reference
Query reference junction type
Specify temperature in °C
Specify temperature in °F
Specify temperature in K
23°C, 73.4°F, 296K
0°C, 32°F, 273K
Query simulated reference
Query default *RST reference
Query lowest allowable reference
Query largest allowable reference