Figure 5-4 Measurement Event Register - Keithley 2010 User Manual

Hide thumbs Also See for 2010:
Table of Contents

Advertisement

Figure 5-4
Measurement event
register
Decimal Weighting
Measurement Event Register:
Bit B0, Reading Overflow (ROF) — Set bit indicates that the reading
exceeds the measurement range of the instrument.
Bit B1, Low Limit1 (LL1) — Set bit indicates that the reading is less than
the Low Limit 1 setting.
Bit B2, High Limit1 (HL1) — Set bit indicates that the reading is greater
than the High Limit 1 setting.
Bit B3, Low Limit 2 (LL2)— Set bit indicates that the reading is less than
the Low Limit 2 setting.
Bit B4, High Limit 2 (HL2)— Set bit indicates that the reading is greater
than the High Limit 2 setting.
Bit B5, Reading Available (RAV) — Set bit indicates that a reading was
taken and processed.
Bit B6 — Not used.
Bit B7, Buffer Available (BAV) — Set bit indicates that there are at least two
readings in the trace buffer.
Bit B8, Buffer Half Full (BHF) — Set bit indicates that the trace buffer is
half full.
Bit B9, Buffer Full (BFL) — Set bit indicates that the trace buffer is full.
Bits B10 through B15 — Not used.
Bit Position
B15 - B12
B11
Event
Value
Value : 1 = Measurement Event Set
0 = Measurement Event Cleared
SCPI Command Reference
B10
B9
B8
B7
B6
B5
BFL
BHF BAV
RAV
32
512
256 128
9
8
7
5
(2 )
(2 )
(2 )
(2 )
0/1
0/1
0/1
0/1
Events : BFL = Buffer Full
B4
B3
B0
B2
B1
HL2
LL2
HL1
LL1
ROF
16
2
8
4
1
4
3
2
1
0
(2 )
(2 )
(2 )
(2 )
(2 )
0/1
0/1
0/1
0/1
0/1
BHF = Buffer Half Full
BAV = Buffer Available
RAV = Reading Available
HL = High Limit
LL = Low Limit
ROF = Reading Overflow