Keithley 2010 User Manual page 220

Hide thumbs Also See for 2010:
Table of Contents

Advertisement

2010 Low-Noise Multimeter
TRUE RMS AC VOLTAGE AND CURRENT CHARACTERISTICS
ACCURACY
1
VOLTAGE
RANGE
100.0000 mV
1.000000
10.00000
100.0000
750.000
CURRENT
RANGE
1.000000
3.00000
HIGH CREST FACTOR ADDITIONAL ERROR
±(% of reading)
Crest Factor:
Additional Uncertainty: 0.05
AC OPERATING CHARACTERISTICS
FUNCTION
ACV (all ranges)
and
ACI (all ranges)
AC SYSTEM SPEEDS
FUNCTION/RANGE CHANGE
AUTORANGE TIME: <3 s.
ASCII READINGS TO RS-232 (19.2K BAUD)
MAX. INTERNAL TRIGGER RATE
MAX. EXTERNAL TRIGGER RATE
ADDITIONAL LOW FREQUENCY ERRORS
±(% of reading)
20 Hz – 30 Hz
30 Hz – 50 Hz
50 Hz – 100 Hz
100 Hz – 200 Hz
200 Hz – 300 Hz
>300 Hz
: ±(% of reading + % of range), 23°C ±5 °C
CALIBRATION
RESOLUTION
CYCLE
0.1 µV
V
1.0 µV
90 Days
V
10 µV
V
100 µV
1 Year
V
1 mV
TEMPERATURE
COEFFICIENT/°C
CALIBRATION
RESOLUTION
CYCLE
A
1 µA
90 Day/1 Year
A
10 µA
90 Day/1 Year
9
TEMPERATURE
COEFFICIENT/°C
7
1–2
2–3
0.15
2
DIGITS
RDGS./s
RATE
6
1
2 3
0.5 (0.4)
SLOW
2 3
6
1
1.4 (1.5)
MED
6
1
2 4
4.0 (4.3)
MED
6
1
2 3
2.2 (2.3)
FAST
2 4
6
1
35 (30)
FAST
2, 5
6
: 4/s.
4
: 50/s.
4
: 300/s.
4
: 300/s.
SLOW
MED
0
0.3
0
0
0
0
0
0
0
0
0
0
3 Hz–
10 Hz–
10
10 Hz
20 kHz
0.35 + 0.03
0.05 + 0.03
0.35 + 0.03
0.06 + 0.03
0.035 + 0.003
0.005 + 0.003
8
3 Hz–
10 Hz–
10 Hz
3 kHz
0.30 + 0.04
0.10 + 0.04
0.35 + 0.06
0.15 + 0.06
0.035 + 0.006
0.015 + 0.006
8
AC GENERAL
INPUT IMPEDANCE: 1MΩ ±2% paralleled by <100pF.
3–4
4–5
ACV INPUT PROTECTION: 1000V.
0.30
0.40
MAXIMUM DCV: 400V on any ACV range.
ACI INPUT PROTECTION: 3A, 250V fuse.
BURDEN VOLTAGE: 1A Range: <0.35V rms.
BANDWIDTH
3 Hz–300 kHz
SHUNT RESISTOR: 0.1Ω on all ACI ranges.
30 Hz–300 kHz
AC CMRR: >70dB with 1kΩ in LO lead.
30 Hz–300 kHz
MAXIMUM CREST FACTOR: 5 at full scale.
300 Hz–300 kHz
VOLT HERTZ PRODUCT: ≤8 × 10
300 Hz–300 kHz
OVERRANGE: 120% of range except on 750V and 3A ranges.
AC NOTES
1. Specifications are for SLOW rate and sinewave inputs >5% of range.
2. Speeds are for 60Hz (50Hz) operation using factory default operating
conditions (*RST). Auto zero off, Auto range off, Display off, includes
measurement and binary data transfer out the GPIB.
3. 0.01% of step settling error. Trigger delay = 400ms.
4. Trigger delay = 0.
5. DETector:BANDwidth 300, NPLC = 0.01.
6. Maximum useful limit with trigger delay = 175ms.
7. Applies to non-sine waves >5Hz and <500Hz. (Guaranteed by design for
Crest Factors >4.3)
FAST
8. Applies to 0°–18°C and 28°–50°C.
9. For signal levels >2.2A add additional 0.4% to "of reading" uncertainty.
10. Typical uncertainties. Typical represents two sigma or 95% of manufac-
1.0
tured units measure < 0.35% of reading and three sigma or 99.7% <
0.18
1.06% of reading.
0.10
0
20 kHz–
50 kHz–
50 kHz
100 kHz
0.11 + 0.05
0.60 + 0.08
0.12 + 0.05
0.60 + 0.08
0.006 + 0.005
0.01 + 0.006
0.03 + 0.01
3 kHz-
5 kHz
0.14 + 0.04
0.18 + 0.06
0.015 + 0.006
3A Range: <1V rms.
7
V·Hz.
100 kHz–
300 kHz
4 + 0.5
4 + 0.5
HW 9/5/03