Error And Status Register - Hach sc1000 User Manual

Hach controller user manual
Hide thumbs Also See for sc1000:
Table of Contents

Advertisement

Standard Operations
Figure 47 Profibus/Modbus configuration menu

5.12.2 Error and status register

Bit
Error
0
Measurement calibration error
1
Electronic adjustment error
2
Cleaning error
3
Measuring module error
4
System re-initialization error
64
7. Select a tag and press the
Select a tag and press the
(Figure 47
and
Table
Table 14 Telegram list—Column description
Column
Description
Profibus: Data position in the configured Profibus slave (in 2 byte words)
Modbus: Data position in the configured Modbus slave
1
This slave contains holding registers beginning at 40001.
Example: "0" means register 40001 or "11" means register 40012.
2
Tag name to identify the configured data.
Data type
float=floating point value
3
int=integer numbers
sel=integer value resulting from a enum (selection) list
Data status
4
r=data are read only
r/w=read/write (for DPV0 only)
8. Repeat the steps to add further devices and tags.
9. Press the
button to store the Profibus configuration.
ENTER
Note: ERROR and STATUS definitions are valid for all sc probes.
Table 15 Error register
button. The new tag is added to the Telegram list.
ENTER
and
button to move the position of the tag
UP
DOWN
14).
Telegram list with new tag
Description
An error has occurred during the last calibration.
An error has occurred during the last electronic
calibration.
The last cleaning cycle failed.
A failure in the measuring module has been detected.
Some settings have been detected inconsistent and set
to factory defaults.