A.5 Attach the external output module
Terminal
1–4
5
6
7
8
9
10
11
12–16
The output module provides two outputs, 0–20 mA/4–20 mA, 500 Ohm.
1. Remove power and disconnect all cable connections to the module.
2. Place the External Output Module on the DIN rail to the right of the base module and
slide firmly against the base module (or other connected modules).
3. Make the appropriate connections as shown in
4. Connect cables to the module and reconnect power from the instrument.
Figure 78 External output module
Table 51 External Output Module Terminal Assignments
Assignment
Not used
Shield
+
Not used
+
Shield
Not used
Figure 78
and
Table
51.
Description
Analog output 1 max. 500 Ohm
Analog output 2 max 500 Ohm