Noisy Output; Dark Patches - Teledyne DALSA HS-40-04k40-xx-R User Manual

High sensitivity line scan ccd camera
Table of Contents

Advertisement

114
Line Dropout, Bright Lines, or Incorrect Frame Rate
Verify that the frequ ency of the internal sync is set correctly, or w hen the cam era is set to
external sync that the EXSYN C signal su pplied to the cam era d oes not exceed the
cam era's u seable fram e rate u nd er the cu rrent operating cond itions.

Noisy Output

Check you r pow er su pply voltage ou tpu ts for noise. N oise present on these lines can
resu lt in poor vid eo qu ality.

Dark Patches

If d ark patches appear in you r ou tpu t the optics p ath m ay have becom e contam inated .
Clean you r lenses and sensor w ind ow s w ith extrem e care.
1. Take stand ard ESD precau tions.
2. Wear latex gloves or finger cots
3. Blow off d u st u sing a filtered blow bottle or d ry, filtered com pressed air.
4. Fold a piece of optical lens cleaning tissu e (approx. 3" x 5") to m ake a squ are pad that
5. Moisten the pad on one ed ge w ith 2-3 d rops of clean solvent—either alcohol or
6. Wipe across the length of the w ind ow in one d irection w ith the m oistened end first,
7. Repeat steps 2-4 u sing a clean tissu e u ntil the entire w ind ow has been cleaned .
8. Blow off any ad hering fibers or particles u sing d ry, filtered com pressed air.
Stuck Bits
If d ata bits seem to be stu ck or d o not change, check that the cam era is not satu rated by
preventing light from entering. To verify the d ata path integrity, check the levels of the 2
test pattern pixels (first 2 pixels follow ing the last End -of-line pixel. You m ay need to tu rn
the End -of-line sequ ence ―on‖ by send ing the com m and
consistent valu e of these pixels (170DN / 85DN ) cou ld be an ind ication of the follow ing:
 shorted bits
 stu ck bits
 d igital noise picku p on the I/ O cable
 open connection.
To activate the test pattern, u se the com m and
proper tim ing and connections betw een the cam era and fram e grabber.
03-032-20013-03
is approxim ately one finger-w id th
acetone. Do not satu rate the entire pad w ith solvent.
follow ed by the rest of the pad . The d ry part of the pad shou ld follow the m oistened
end . The goal is to prevent solvent from evaporating from the w ind ow su rface, as this
w ill end u p leaving resid u e and streaking behind .