5.7 Save Calibration Data and Exit

Proceed as follows to save the calibration data, and to exit the Maintenance mode:
1. Remove all test leads from the test tool inputs.
2. Press
Calibration data are valid
Save data and EXIT maintenance?
Calibration data valid indicates that the calibration adjustment procedure
is performed correctly. It does not indicate that the test tool meets the
characteristics listed in Chapter 2.
3. Press
The calibration number and date will be updated only if the calibration
data have been changed and the data are valid.
The calibration data will change when a calibration adjustment has been
done. The data will not change when just entering and then leaving the
maintenance mode without doing a calibration adjustment.
The calibration number and date will NOT be updated if only the display
contrast has been adjusted.
Possible error messages.
Invalid calibration data:
WARNING.Calibration data NOT valid.
Save data and EXIT?
Proceed as follows:
To return to the Maintenance mode:
Press
Now press
test tool, starting at Section 5.5.
to start the calibration.
Cap. Gain (CL 0960):READY
(EXIT). The test tool will display:
(YES) to save and exit.
NO.
until the display shows
.
Note
Notes
WarmingUp (CL 0200):IDLE
Calibration Adjustment