5.6.9 Calculate Gain

1. Remove all test leads from the test tool inputs.
2. Press
3. Press
4. Wait until the display shows the calibration status
5. Continue at section 5.7

5.7 Save Calibration Data and Exit

Proceed as follows to save the calibration data, and to exit the Maintenance mode:
1. Remove all test leads from the test tool inputs.
2. Press
Calibration data valid.
Save data and exit maintenance mode?
Calibration data valid indicates that the calibration adjustment procedure
is performed correctly. It does not necessarily mean that the test tool meets
the characteristics listed in Chapter 2.
3. Press
After saving the calibration data, the calibration number and - date will be
updated if the calibration data have been changed and the data are valid.
The calibration number and - date will not change if:
- the calibration mode is entered and left without doing a calibration adjustment.
- only the contrast calibration adjustment (5.4) and/or the probe calibration is
done.
If you press
either calibrate the test tool again, or press
exit.
Table 5-6. Ohm Gain Calibration Points
to select calibration adjustment step CL 0920.
to start the calibration.
. The test tool will display:
EXIT
Note
to save and exit.
YES
Note 1
Note 2
,the test tool returns to the calibration mode. You can
NO
Calibration Adjustment
UUT input Value (5500 NORMAL)
100 Ω
1 kΩ
10 kΩ
100 kΩ
1 MΩ
10 MΩ
.
:READY
,
to save and
EXIT
YES
5

5.6.9 Calculate Gain