Fluke 43B Service Manual page 75

Power quality analyzer
Hide thumbs Also See for 43B:
Table of Contents

Advertisement

Figure 4-6. Test Tool Input 1-B to 5500A Normal Output for >300V
2. Select the AUTO test tool setup:
Press
to select the MENU.
Press
till SCOPE is highlighted.
Press
to select SCOPE mode
3. Select DC coupling & reading for Input 1 and 2.
Press
to select menu SCOPE SETUP.
Press
to select Input 1 READING.
Press
to highlight DC.
Press
to confirm;
Press
to highlight Input 1 Coupling.
Press
to select the Input 1 Coupling menu.
Press
to highlight DC Coupling.
Press
to confirm;
Press
to select Input 2 READING.
Press
to select the Input 2 READING
Press
to highlight DC.
Press
to confirm;
Press
to highlight Input 2 Coupling.
Press
to select the Input 2 COUPLING.
Press
to highlight DC Coupling.
Press
to confirm;
Press
to return to SCOPE.
4. Select the appropriate sensitivity for the test tool:
4.5 Input 1 and Input 2 Tests in the SCOPE MODE.
mark changes to I.
mark changes to I.
mark changes to I.
mark changes to I.
Performance Verification
ST8129.WMF
4
4-23