Contacting a Service Center................1-5 Safety Information: Read First................1-5 About This Manual ..................2-1 Introduction......................2-1 Users Manual Contents ..................2-1 Features Of Fluke 433/434................3-1 Introduction......................3-1 General Measurements ..................3-1 Measuring modes to investigate details .............3-2 Basic Operations and Menu Navigation ............4-1 Introduction......................4-1...
Page 4
Fluke 433/434 Users Manual Screen Types......................5-2 Screen information common for all screen types..........5-2 Input Connections................... 6-1 Introduction......................6-1 Input Connections ....................6-1 Scope Waveform and Phasor ................ 7-1 Introduction......................7-1 Scope Waveform....................7-1 Scope Phasor......................7-2 Tips and Hints....................7-2 Volts/Amps/Hertz .................... 8-1 Introduction......................8-1 Table ........................8-1 Trend........................8-2...
Page 5
Tips and Maintenance..................20-1 Introduction......................20-1 Cleaning the Analyzer and its Accessories ............20-1 Storing the Analyzer ..................20-1 Keeping the Battery in Good Condition ............20-1 Installation of Options in Fluke 433 ..............20-1 Parts and Accessories..................20-2 Troubleshooting ....................20-3 Specifications....................21-1 Introduction......................21-1 Electrical Measurements..................21-2 Trend Recording ....................21-11...
Page 6
Fluke 433/434 Users Manual Memory......................21-13 Printers and Interface ..................21-14 Power Supply and Battery Charger..............21-14 Mechanical......................21-15 Environmental....................21-15 Electro Magnetic Compatibility (EMC) ............21-16 Safety .........................21-16...
Chapter 1 General Aspects Introduction This chapter informs you about a number of general and important aspects concerning the Fluke 433/434 Three Phase Power Quality Analyzer (hereafter referred to as ‘Analyzer’). This concerns: • Warranty and Liability Conditions. • Declaration of Conformity.
Fluke's warranty obligation is limited, at Fluke's option, to refund of the purchase price, free of charge repair, or replacement of a defective product which is returned to a Fluke authorized service center within the warranty period.
General Aspects Declaration of Conformity Declaration of Conformity Declaration of Conformity Fluke 433/434 Three Phase Power Quality Analyzers Manufacturer Fluke Industrial B.V. Lelyweg 1 7602 EA Almelo The Netherlands Statement of Conformity Based on test results using appropriate standards, the product is in conformity with...
Fluke 433/434 Users Manual Shipment Note The following items are included in your Analyzer Kit: Note: When new, the Analyzer’s rechargeable NiMH battery is not charged. Refer to Chapter 4 – Powering the Analyzer. (5x) Fluke 434 (5x) (4x) Figure 1-1. Contents of Analyzer Kit...
Optical Cable for USB, standard for Fluke 434. Optional for Fluke 433 Contacting a Service Center To locate a Fluke authorized service center, visit us on the World Wide Web at: www.fluke.com or call Fluke using any of the phone numbers listed below: +1-888-993-5853 in the U.S.
Page 12
• Use only insulated current probes, test leads and adapters as supplied with the Analyzer, or indicated as suitable for the Fluke 433/434 Analyzer. • Before use, inspect the Analyzer, voltage probes, test leads and accessories for mechanical damage and replace when damaged.
Page 13
General Aspects Safety Information: Read First Slider switch on BC430 Battery Charger / Power Adapter to select line power voltage: 115V 230V Max. Input Voltage at Voltage Banana Inputs to Ground: Input A (L1), B (L2), C (L3), N to Ground: 1000 V Cat III, 600 V Cat IV. Max.
This Users Manual gives full and comprehensive information on how to use the Fluke 433 and 434 Three Phase Power Quality Analyzers effectively and in a safe manner. Read it carefully to learn about safe use of the Analyzer and its accessories and to take full advantage of all measuring modes.
Page 16
Fluke 433/434 Users Manual • Chapter 18. Setting up the Analyzer: a comprehensive explanation of adjustments to customize measurements. • Chapter 19. Using Memory, printer and PC: how to save, recall and delete screenshots and data formats. How to make hard copies of measurement results and setup of communication with PC.
Fluke 434 has additional features such as Interharmonics, Transients, Energy Usage, extra memory to store Screens and Data, FlukeView software, and an optical isolated interface cable. In Fluke 433 these functions can be installed optionally. If not installed, they are shown in the menus in grey color.
Fluke 433/434 Users Manual Measuring modes to investigate details Phase voltages. Should be close to the nominal value. Voltage waveforms must be a sine wave that is smooth and free from distortion. Use Scope Waveform to check the waveform shape. Use Dips & Swells to record sudden voltage changes. Use Transients mode to capture voltage anomalies.
Chapter 4 Basic Operations and Menu Navigation Introduction This chapter deals with a number of general aspects of the Analyzer’s operation: • Tilt Stand and Hang Strap • Powering the Analyzer • Display Brightness • Locking the keyboard • Menu navigation •...
Fluke 433/434 Users Manual A hang strap is supplied with the Analyzer. The figure below shows how to attach the strap correctly to the Analyzer. Figure 4-2. Fixing the hang strap Powering the Analyzer The Analyzer has a built-in rechargeable NiMH battery that can power it for more than 6 hours when fully charged.
Basic Operations and Menu Navigation Display Brightness Caution To prevent decrease of battery capacity, charge it at least twice a year. Power On/Off: Press to power up or down with the last setup configuration. The welcome screen shows what Analyzer settings are currently in use.
Fluke 433/434 Users Manual Menu Navigation Most of the Analyzer functions are menu operated. Arrow keys are used to navigate through menus. The Function keys F1 ... F5 and the ENTER key are used to make selections. Active Function key selections are highlighted with a black background.
Chapter 5 Display Information Introduction The Analyzer uses five different screen types to present measuring results in the most effective way. The features these screens have in common are explained in this chapter. Details that are specific for a certain measuring mode are presented in the chapter explaining that mode.
Fluke 433/434 Users Manual Phase Colors Measuring results belonging to different phases are presented with individual colors. If for a certain phase voltage and current are displayed simultaneously, the voltage color has a dark tone and the current has a light tone. The set of phase colors can be chosen via the SETUP key and function key F4 –...
Page 25
Display Information Screen information common for all screen types Status indicators. The following symbols may appear on the screen to show the state of Analyzer and measurements: Time that a measurement has been going on. Format: hours, minutes, seconds. When waiting for a timed start, time counts down with prefix -.
Chapter 6 Input Connections Introduction This chapter explains how to make connection to the power distribution system under test and how to adjust the Analyzer settings. Check that the Analyzer setup meets the characteristics of the system under test and the accessories that are used.
Page 28
Fluke 433/434 Users Manual Figure 6-1. Mounting the decals for voltage and current inputs De-energize power systems before making connections whenever possible. Avoid working alone and work according to the warnings listed in Chapter 1, Safety Information. For a 3-phase system make the connections as shown in Figure 6-2.
Page 29
Input Connections Input Connections Before making any measurements, set the Analyzer up for the line voltage, frequency, and wiring configuration of the power system you want to measure. This is explained in Chapter 18, General Settings. Scope Waveform and Phasor display are useful to check if voltage leads and current clamps are connected correctly.
Chapter 7 Scope Waveform and Phasor Introduction Scope mode shows voltages and currents in the power system under test by means of waveforms or vector diagram. Also numerical values are shown such as phase voltages, phase currents, frequency, and phase angles between voltages and currents. Scope Waveform To access the Scope Waveform screen: SCOPE...
Fluke 433/434 Users Manual PERSISTence ON memorizes all waveform variations on screen. Switch between HOLD and RUN of screen update. Cursor. When the Cursor is on, the waveform values at the Cursor are displayed in the screen header. Positioning the Cursor across the left or right screen end brings the next screen out of a maximum of 6 within viewing area.
Chapter 8 Volts/Amps/Hertz Introduction Volts/Amps/Hertz displays a table with important numerical measuring values. The related Trend screen shows the changes over time of all values in the table. Table To access the VOLTS/AMPS/HERTZ Table screen: MENU ENTER The table gives an overview of voltages and currents in all phases. Also frequency and Crest Factors are shown.
Fluke 433/434 Users Manual Available function keys: Access to the Trend screen. For description see below. Switch between HOLD and RUN of screen update. Switching from HOLD to RUN invokes a menu to select immediate (NOW) or TIMED start time which allows you to define start and duration of the measurement.
Volts/Amps/Hertz Tips and Hints Tips and Hints Voltage and frequency should be close to the nominal values of for example 120 V, 230 V, 480 V, 60 Hz, or 50 Hz. The voltages and currents in the table can e.g. be used to check if power applied to a 3- phase induction motor is in balance.
Chapter 9 Dips & Swells Introduction Dips & Swells records Dips, Interruptions, Rapid Voltage Changes, and Swells. Dips (Sags) and Swells are fast deviations from the normal voltage. Magnitude may be ten up to hundreds of volts. Duration may vary from a half cycle to a few seconds as defined in EN61000-4-30.
Page 38
Fluke 433/434 Users Manual Figure 9-2. Characteristics of a voltage swell During an Interruption the voltage sinks well below its nominal value. In three phase systems an interruption begins when the voltage on all phases are below threshold and ends when one phase is equal to or above the interruption threshold plus hysteresis. The trigger conditions for interruptions are threshold and hysteresis.
Dips & Swells Trend VOLTAGE TOLERANCE RAPID VOLTAGE CHANGE - CHG- VOLTAGE TOLERANCE STEADY TIME VOLTAGE STEP VOLTAGE STEP TRANSITION TIME STEADY TIME TRANSITION TIME TIME Figure 9-4. Characteristics of a rapid voltage change In addition to the voltage, current is also recorded. This allows you to see cause and effect of deviations.
Fluke 433/434 Users Manual The screen builds up from the right side of the screen and the corresponding values are displayed in the screen header. Available function keys: Assign up/down arrow keys to select the voltage or current channels to be displayed.
Dips & Swells Tips and Hints The following Abbreviations and Symbols are used in the tables: Abbreviation Description Symbol Description Rapid Voltage Change Rising voltage edge Voltage Dip Falling voltage edge Voltage Interruption Voltage Swell Available function keys: Switch between NORMAL and DETAILED event table. Return to Trend screen.
Chapter 10 Harmonics Introduction Harmonics measures and records harmonics and interharmonics (Fluke 434) up to the . Related data such as DC components, THD (Total Harmonic Distortion), and K- factor are measured. Harmonics are periodic distortions of voltage, current, or power sinewaves.
Page 44
Selection of waveform set to be used: A (L1), B (L2), C (L3), N (neutral) or ALL Access the table screen. Interharmonics display on/off (Fluke 434 only). Switch between HOLD and RUN of screen update. Switching from HOLD to RUN invokes a menu to select immediate (NOW) or TIMED start time which allows you to define start and duration of the measurement.
Harmonics Table Table To access the Harmonics Table screen: The table display shows 8 measurements per phase. Using the SETUP key and function key F3 - FUNCTION PREF you can choose the table contents. For detailed information see Chapter 18, FUNCTION PREFerences. Available function keys: Return to Bar Graph screen.
Fluke 433/434 Users Manual Assign arrow keys to Cursor or vertical Zoom operation. Return to table screen. Switch between HOLD and RUN of screen update. Switching from HOLD to RUN invokes a menu to select immediate (NOW) or TIMED start time which allows you to define start and duration of the measurement.
Power & Energy displays a table with all important power parameters. The related Trend screen shows the changes over time of all measuring values in the table. Fluke 434 can also display energy usage and offers verification of energy meters with a pulse contact. For power calculations you can choose Fundamental or Full.
Page 48
Symbols indicate if a load is capacitive ( ) or inductive ( ). A popup table with energy usage by phase and in total can be activated on the Fluke 434 by pressing the F3 – ENERGY softkey. The table shows real energy (kWh), apparent energy (kVAh), and reactive energy (kVARh) The energy measurement starts when Power &...
Power & Energy Trend To access the Energy popup table: Available function keys: Pulse count mode on/off. Switch Energy screen on/off. Access Trend screen. For description see below. Switch between HOLD and RUN of screen update. Switching from HOLD to RUN invokes a menu to select immediate (NOW) or TIMED start time which allows you to define start and duration of the measurement.
Fluke 433/434 Users Manual PREFerences. With the demand interval set to OFF the Trend functions as usual with automatic horizontal scaling. Available function keys: Assign up/down arrow keys to select a row from the table for Trend display. The selected row is displayed in the screen header.
Chapter 12 Flicker Introduction Flicker quantifies the luminance fluctuation of lamps caused by supply voltage variations. The algorithm behind the measurement meets EN61000-4-15 and is based on a perceptual model of the human eye / brain sensory system. The Analyzer converts duration and magnitude of voltage variations into an ‘annoyance factor’...
Page 52
Fluke 433/434 Users Manual ENTER Flicker is characterized by: short term severity Pst (measured over 1 min for fast feedback), short term severity Pst (measured over 10 min) and a long term severity Plt (measured over 2 hours). This data and also the related D-parameters Dc, Dmax, and TD (acc.
Flicker Trend Trend To access Flicker Trend screen: The parameters in the table update over time. They are recorded whenever the measurement is on. Trend displays the changes in these values over time. All values in the table are recorded, but the Trends from each row in the table are displayed one at a time.
Fluke 433/434 Users Manual Tips and Hints Use the PF5 flicker trend and half-cycle voltage or current trends to find the source of flicker. Press function key F1 to assign the arrow keys to selection of flicker, voltage, and current trends.
Chapter 13 Unbalance Introduction Unbalance displays phase relations between voltages and currents. Measuring results are based upon the fundamental frequency component (60 or 50 Hz). In a 3-phase power system, the phase shift between voltages and between currents should be close to 120°. Unbalance mode offers a measurement table, a related Trend display, and a Phasor display.
Fluke 433/434 Users Manual The table screen shows all relevant numerical values: negative voltage unbalance percentage, zero sequence voltage unbalance percentage (in 4-wire systems), negative current unbalance percentage, zero sequence current unbalance percentage (in 4-wire systems), fundamental phase voltage, frequency, fundamental phase current, angle between phase-neutral voltages relative to the reference phase A/L1 and angles between voltage and current for each phase.
Unbalance Phasor Cursor. When the Cursor is on, the Trend values at the Cursor are displayed in the screen header. Moving the Cursor off the left or right side of the screen brings the next of six screens into the viewing area. Zoom.
Page 58
Fluke 433/434 Users Manual unbalance should not exceed 10 %. In case of too high unbalance, use other measuring modes to further analyze the power system. Each phase voltage or current can be split-up into three components: positive sequence, negative sequence, and zero sequence.
Transients Introduction The Fluke 434 Analyzer can capture waveforms at high-resolution during a variety of disturbances. The Analyzer will give a snapshot of the voltage and current waveforms at the precise time of the disturbance. This allows you to see the waveforms during dips, swells, interruptions, current swells and transients.
Page 60
Fluke 433/434 Users Manual ENTER In the Start menu you can choose a trigger event or a combination of trigger events, transients (Volt) and current (AMP) trigger level, and Immediate or Timed start of the measurement. The Analyzer may be set up to capture waveforms each time it sees: Voltage Transient, Voltage Swell, Voltage Dip, Voltage Interruption, or Current swell.
Transients Tips and Hints Assign up/down arrow keys to browse through all captured screens. Persistence on memorizes all waveform variations. Switch between HOLD and RUN of screen update. Switching from HOLD to RUN invokes a menu to select immediate (NOW) or TIMED start time which allows you to define start and duration of the measurement.
Inrush Currents Introduction Inrush Currents can be captured by Fluke 434. Inrush Currents are surge currents that occur when a large, or low-impedance load comes on line. Normally the current will stabilize after some time when the load has reached normal working condition. For example the start-up current in induction motors can be ten times the normal working current.
Page 64
Fluke 433/434 Users Manual ENTER Use the arrow keys in the Start menu to adjust the trigger limits: expected inrush time, maximum current, nominal current, threshold, and hysteresis. The maximum current determines the vertical height of the current display windows. Threshold is the current level that triggers the trend capture.
Page 65
Inrush Currents Inrush Trend Display MAXIMUM AMPS DURATION THRESHOLD (= TRIGGER) HYSTERESIS zero pretrigger inrush time 35 s 12.3 A rms during inrush 12.3 A 1.5 A Figure 15-1. Inrush characteristics and relation with start menu Use Cursor and Zoom to investigate details of the recorded Trends. Selection of channels to be displayed is done with the up/down arrow keys.
Fluke 433/434 Users Manual Tips and Hints Check the peak currents and their duration. Use the Cursor for readout of momentary values. Check if fuses, circuit breakers, and conductors in the power distribution system can withstand the inrush current during this period. Check also if phase voltages stay stable enough.
Chapter 16 Power Quality Monitoring Introduction Power Quality Monitoring or System Monitor displays a Bar graph screen. This screen shows whether important Power Quality parameters meet requirements. Parameters include: 1. RMS voltages 2. Harmonics 3. Flicker 4. Dips/Interruptions/Rapid Voltage Changes/Swells 5.
Page 68
Fluke 433/434 Users Manual Figure 16-1. Power Quality Monitor Main Screen The length of a bar increases if the related parameter is further away from its nominal value. The bar turns from green to red if an allowed tolerance requirement is violated.
Page 69
Power Quality Monitoring Introduction average from these measuring values across 10-minute observation periods. The 10- minute averages are compared against the tolerance range (in this example 102 ... 138 V). The 100 % limit means that the 10-minute averages must always (i.e. 100 % of time or with 100 % probability) be within range.
Fluke 433/434 Users Manual Power Quality Main Screen To access the Power Quality Main screen: MONITOR Power Quality Monitoring is reached via the MONITOR key and a menu for Immediate or Timed start. With the left/right arrow keys you can position the Cursor on a particular Bar Graph.
Power Quality Monitoring Events Table Events Table Figure 16-2. Events Table The events table shows the events that occurred during the measurement with date/time of start, phase, and duration. The amount of information in the table can be selected with the Function keys F2 and F3: •...
Fluke 433/434 Users Manual Two ways to access Trend: 1. Use the up/down arrow keys to highlight an event in the table. To access Trend press the ENTER key. The Cursor is on, in the mid of screen and located on the selected event.
Power Quality Monitoring Bar Graph Screen Bar Graph Screen Figure 16-4. Bar Graph Screen The main system monitor display shows the worst harmonic for each of the three phases. Function key F2 brings up a screen with Bar Graphs showing the percentage of time each phase spent within limits for 25 harmonics and Total Harmonic Distortion (THD).
Chapter 17 Cursor and Zoom Introduction This chapter explains how to use Cursor and Zoom to display and investigate details of Waveform, Trend, and Bar Graph displays. Cursor and Zoom have a certain amount of interaction and are both operated by the arrow keys. The Cursor is a vertical line that can be positioned on a point on a Waveform, Trend, or Bar Graph.
Fluke 433/434 Users Manual Figure 17-3. Waveform display with cursor and zoom on Press Function key F2 to obtain a subset with keys to control Cursor and Zoom: • Press F3 to switch the Cursor on. Use the left/right arrow keys to move the Cursor horizontally along the waveforms.
Cursor and Zoom From Events Table to Trend Display with Cursor On. Figure 17-6. Trend display with cursor and zoom on The Function keys F1, F2, and F3 and the arrow keys are used to operate Cursor and Zoom: • Operate F2 to switch the Cursor on.
Fluke 433/434 Users Manual Press F5 to obtain trend display with the cursor on and positioned on the highlighted event in the table. Cursor on Bar graph Displays As an example the Three-phase Voltage Harmonics display as shown in Figure 17.7 is used.
Chapter 18 Setting up the Analyzer Introduction The SETUP key accesses menus to view and change Analyzer settings. At delivery the Analyzer is adjusted to settings that match your local situation and the supplied accessories. The table below gives an overview: Setting Preset Value Nominal Voltage...
Page 80
Fluke 433/434 Users Manual Figure 18-1. Welcome screen at power-on The Settings are grouped in four functional sections and are explained accordingly in four sections of this manual chapter: • General Settings: Date, Time, wiring Configuration, nominal Voltage, nominal Frequency, current and voltage probe type, information language, survey and installation of options.
Setting up the Analyzer General Settings General Settings To access the General Settings menus: SETUP SETUP ENTER The actual settings are shown in the SETUP entry screen. Use the key operations described above to change an item. Read below how to make adjustments: User name/address: see section USER PREFerences.
Page 82
Fluke 433/434 Users Manual Freq: adjustment of Nominal Frequency. Use the up/down arrow keys to select 60 or 50 Hz. Press Function key F5 – OK to confirm. Limits: see section Limits Settings. Clamp, A range, V scale: adjustment of the Analyzer to the characteristics of current clamps and voltage leads.
Page 83
Setting up the Analyzer General Settings Below you will find a step-by-step example on how to change wiring configuration to 3- phase Wye IT (IT = Interrupted Terra = Interrupted Ground). SETUP The active configuration is indicated behind Config. Config is highlighted indicating that this item can be adjusted when you press the ENTER key.
Page 84
Fluke 433/434 Users Manual An information screen shows up that informs you on how to connect the Analyzer to the power system under test. When done press F5. Return to Setup entry screen. The new configuration is indicated behind Config. and the belonging configuration symbol is shown on the right side of the screen.
Setting up the Analyzer FUNCTION PREFerences FUNCTION PREFerences To access the FUNCTION PREFerences menus: SETUP SETUP ENTER FUNCTION PREFerences allows you to customize data presentation of measuring functions. This concerns for instance Offset and Span of Trend and Waveform displays. The table below gives a survey adjustable items for each function.
Page 86
Fluke 433/434 Users Manual Measuring Function/ Measuring Data to be adjusted Settings Type Screen Type 1. Scope Waveform Volt, Amp (separate for Phase Range and Neutral) 2. Volts/Amps/Hertz Trend Volt (Peak), Amp (Peak), CF, Offset + Span (separate for Phase and Neutral),...
Page 87
Setting up the Analyzer FUNCTION PREFerences The example below shows stepwise how to adjust offset and span of a Volts/Amps/Hertz trend after a voltage change has occurred. Trends are outside their window. SETUP Press SETUP to access setup entry screen. Press function key F3 to access select function screen.
Page 88
Fluke 433/434 Users Manual Use the left/right arrow keys to decrease the Voltage offset. Use the up/down arrow keys to select Volt span adjustment. Use the left/right arrow keys to increase the Voltage span. Press function key F5 three times to return to the Volts/Amps/Hertz trend screen with new offset and span.
Setting up the Analyzer USER PREFerences USER PREFerences To access the USER PREFerences menus: SETUP SETUP ENTER USER PREFerences allows you to customize Phase Identification and Colors, Printer and RS-232 settings, Auto shut-off, definition of User name/address (as shown in entry screen), and display contrast.
Page 90
Fluke 433/434 Users Manual Phase Colors: use the function keys F1 ... F4 to choose colors as used in the USA, EU, UK, or according to IEC. Or define your own set of colors: use the up/down arrow keys to select a phase and use the left/right arrow keys to select a color.
Setting up the Analyzer Limits Adjustments Limits Adjustments To navigate the Limits Setup menus: SETUP SETUP ENTER 0...9 A...Z 18-13...
Page 92
Fluke 433/434 Users Manual Limits Adjustments is used to save, recall, and define sets of limits for: • Power Quality Monitoring. • Dips/Interruptions/Rapid Voltage Changes/Swells. Read below how to do this: Adjust Monitor Limits is the entry menu. It shows the main settings of the active set of limits: name, creation date, and a summary of limits data.
Page 93
Setting up the Analyzer Limits Adjustments View Monitor Limits menu. This menu has the same structure as the Edit Monitor Limits menu and can be used to view limits without the risk of changing them. Press Function key F5 – BACK to return to the SETUP entry menu. Setup of Monitor Limits, a survey of adjustments.
This allows you to view and analyze all screens belonging to the measurement, and to use Cursor and Zoom. A maximum of 10 datasets can be saved in Fluke 434 and 5 in Fluke 433. Symbol for datasets: Making a Screenshot SAVE Press this key to make a screenshot.
Page 96
Fluke 433/434 Users Manual Memory Operations The MEMORY button accesses menus to save, recall, view, delete and print datasets and screenshots. When you press the MEMORY button, the current measurement screen is frozen. MEMORY Available function keys (in the sequence they are normally used): SAVE.
PC or laptop. The information supplied with FlukeView software informs you about its features. The interface connection is located at the right Analyzer side and attainable if the tilt stand is folded out. For Fluke 433 the interface cable and FlukeView software can be ordered as an option.
Page 98
Fluke 433/434 Users Manual the up/down arrow keys and ENTER. Then adjust the baudrate with the left/right arrow keys and leave the menu with F5 - BACK. Baudrate and COM port number in FlukeView must be adjusted correctly. Figure 19-2. Analyzer and laptop PC For correct communication with a printer it is necessary that baudrate and printer type of Analyzer match with the hard copy device.
The Advanced Functions Interharmonics, Transients, Energy Usage, Inrush Currents, and extra memory as available in Fluke 434, can be activated in an existing Fluke 433. Activation can be done by the user done via a pin-code that is unique for the serial number of your Analyzer.
Press ENTER to confirm the selection and to activate the option. The menu now will show INSTALLED behind the option just activated. For your Fluke 433 you can also order an Upgrade Kit. The kit includes access to install the Advanced Functions and also FlukeView software and an optical interface cable.
Advanced Functions for Fluke 433 (Interharmonics, Transients, Energy Usage, Fluke-433/AF Inrush Currents, Extra Memory). FlukeView Software for Fluke 433 SW43W (V3.0) Upgrade Kit for Fluke 433 (Advanced Functions, FlukeView Software, Optical Fluke-433/UGK Cable for USB model OC4USB) Optical Isolated RS-232 Cable PM9080...
Page 102
Fluke 433/434 Users Manual Screen remains black. Make shure that the Analyzer is on: at power-on you should hear a double beep. If the screen remains black, you might have a problem with the screen contrast. Proceed as follows to change Contrast: •...
Specifications Introduction Performance Characteristics Fluke guarantees the properties expressed in numerical values within the tolerances stated. Numerical values without tolerances are typical and represent the characteristics of an average instrument excluding accessories. The Analyzer meets the specified accuracy 30 minutes and two complete acquisitions after power-on. All operational specifications are valid under the restrictions mentioned in section ‘Environmental’...
Fluke 433/434 Users Manual Electrical Measurements The following specifications of the instrument are verified using the “implementation verification” table 3 as specified in 61000-4-30 chap-6-2. FREQUENCY MEASUREMENT Selected Nominal Measurement Range Resolution Accuracy Frequency (Fnom) 50 Hz 42.50 ... 57.50 Hz 0.01 Hz...
Page 105
Specifications Electrical Measurements RMS Voltage Selected Nominal Measurement Range Resolution Accuracy Voltage (Vnom) (CF ≤ 1.4 at full scale) 60 ... 125 Vrms 1.0 ... 250.0 Vrms 0.1 Vrms ± 0.5 % of Vnom 125 ... 250 Vrms 1.0 ... 500.0 Vrms 0.1 Vrms ±...
Page 106
Fluke 433/434 Users Manual Measurement Measurement Range Resolution Accuracy Relative Amplitude 0.0 ... 100.0 % 0.1 % ± 0.1 % ± n x 0.1 %, (±0.4% for %r) Absolute Amplitude Vnom: 60 ... 125 Vrms 0.0 ... 250.0 Vrms ± 5 % of meas ± 2 0.1 Vrms...
Page 107
Specifications Electrical Measurements Voltage Interrupts Settings Adjustment Range Resolution Additional Information Interrupt Threshold level 0.0 ... 50.0 % of Vnom 0.1 % results based on ½ cycle Interrupt Hysteresis level 0.0 ... 10.0 % of Vnom 0.1 % Measurements Measurement Range Resolution Accuracy Interrupt Magnitude...
Page 108
Fluke 433/434 Users Manual Measurements Measurement Range Resolution Accuracy RMS voltage in Transient 10 ... 1000 Vrms 1 Vrms ± 2.5 % of Vnom Function Transient voltage 0 ... ± 6000 Vpeak ± 15 % of meas (TTRANS > 10 µs)
Page 110
Fluke 433/434 Users Manual Inrush Current Settings Adjustment Range Resolution Additional Information Inrush Threshold level 0 ... 999 % of Inom Results based on Irms ( Itrh - Ihys > Inom ) Inrush Hysteresis level 0 ... 999 % of Inom Inrush Evaluation time 7.5 s, 15 s, 30 s, 1.5 m,...
Fluke 433/434 Users Manual Printers and Interface Item Specification Additional Information To be used with interface cable to Type RS-232, optically isolated RS-232 with 9-pole D-plug male ( PM9080) or USB (OC4USB) Spacing “0” Light “1” No light Baud rate 1200, 2400, 9600 …...
Electro Magnetic Compatibility (EMC) Item Customer Specification Additional Information Emission and Immunity EN-61326 Fluke 433/434, including standard accessories, conforms with the EEC directive 89/336 for EMC immunity, as defined by EN- 61326, with the addition of the table below Frequency Disturbance <...
Need help?
Do you have a question about the 433 and is the answer not in the manual?
Questions and answers