Save Calibration Data And Exit - Fluke 435-II Service Manual

Three phase energy and power quality analyzer
Hide thumbs Also See for 435-II:
Table of Contents

Advertisement

Fluke 434-II/435-II/437-II
Service Manual
Cal step
CL 0410
CL 0420
CL 0430
CL 0440
9. Continue at section 5.7

5.7 Save Calibration Data And Exit

Proceed as follows to save the calibration data, and to exit the Maintenance mode:
1. Remove all test leads from the Analyzer inputs.
2. Press
3. Press
After saving the calibration data, the calibration number and - date will be
updated if the calibration data have been changed and the data are valid.
The calibration number and - date will not change if:
- the calibration mode is entered and left without doing a calibration adjustment.
- only the contrast calibration adjustment (5.4) and/or the probe calibration is
done.
If you press
either calibrate the Analyzer again, or press
and exit.
Possible error messages.
The following messages can be shown on the Analyzer display:
5-12
Table 5-2. Voltage Gain Adjustment
Gain125Volt
Gain250Volt
Gain500Volt
Gain6KVolt
---
---
F5
. The Analyzer will display:
EXIT
Calibration data valid.
Save data and exit maintenance mode?
Calibration data valid indicates that the calibration adjustment procedure
is performed correctly. It does not necessarily mean that the Analyzer
meets the characteristics listed in the Analyzer's Users Manual (Refer to
Chapter 2).
F5
to save and exit.
YES
F4
,the Analyzer returns to the calibration mode. You can
NO
WARNING: Calibration data not valid.
Save data and exit maintenance mode?
Description
Note
Note 1
Note 2
F5
Calibrator Setting
120 V, 50 Hz, OPR
230 V, 50 Hz, OPR
480 V, 50 Hz, OPR
600 V, 50 Hz, OPR
STBY
F5
,
to save
EXIT
YES