Transients - Fluke 43B Service Manual

Power quality analyzer
Hide thumbs Also See for 43B:
Table of Contents

Advertisement

43B
Service Manual

4.12 Transients.

Proceed as follows to test the TRANSIENTS function:
1. Connect the test tool to the 5500A as show in Figure 4-12.
2. Press
3. Press
4. Press
VOLTAGE CHANGE will be highlighted, continue with:
1. Press
2. Press
3. Press
4. Press
5. Set the 5500A to 20V, 60Hz (NORMAL output, WAVE sine).
6. Press
7. Set the 5500A to 22.5V.
8. Now transients should be captured. After 40 transients (maximum) the 40th transient
9. When you are finished, set the 5500A to Standby.
4-32
Figure 4-12. Test Tool Input 1 to 5500A Normal Output
.
to highlight TRANSIENTS.
to select the TRANSIENTS mode.
.
to select 20%.
to confirm;
to select START.
to start the test. No transients should be captured.
will be visible.
mark changes to I.
ST8586.WMF