43B
Service Manual
3. Select DC coupling & reading for Input 1 and 2.
4. Set the 5500A to source the appropriate DC voltage from table 4-2.
5. Observe the main reading and check to see if it is within the range shown under the
6. Select the appropriate sensitivity for the test tool:
7. Continue through the test points.
8. When you are finished, set the 5500A to 0 (zero) Volt, and to Standby.
4-16
•
Press
to select menu SCOPE SETUP.
•
Press
to select Input 1 READING.
•
Press
to highlight DC.
•
Press
to confirm;
•
Press
to highlight Input 1 Coupling.
•
Press
to select the Input 1 Coupling menu.
•
Press
highlight DC Coupling.
to
•
Press
to confirm;
•
Press
to select Input 2 READING.
•
Press
to select the Input 2 READING
•
Press
to highlight DC.
•
Press
to confirm;
•
Press
to highlight Input 2 Coupling.
•
Press
to select the Input 2 COUPLING.
•
Press
to highlight DC Coupling.
•
Press
to confirm;
•
Press
to return to SCOPE.
appropriate column.
•
Press
to select RANGE 1 or RANGE 2.
•
Press
to select RANGE 1 or press
•
Press
to select the ranges mentioned in the table.
mark changes to I.
mark changes to I.
mark changes to I.
mark changes to I.
to select RANGE 2.