43B
Service Manual
2. Select the AUTO test tool setup:
3. Select DC coupling & reading for Input 1 and 2.
4. Select the appropriate sensitivity for the test tool:
5. Select the appropriate timebase setting for the test tool
6. Set the 5500A to source the appropriate AC voltage.
7. Observe the Input 1 and Input 2 main reading and check to see if it is within the
8. Continue through the test points.
9. When you are finished, set the 5500A to Standby.
4-18
•
Press
to select the MENU.
•
Press
till SCOPE is highlighted.
•
Press
to select SCOPE mode
•
Press
to select menu SCOPE SETUP.
•
Press
to select Input 1 READING.
•
Press
to highlight ACrms.
•
Press
to confirm;
•
Press
to highlight Input 1 Coupling.
•
Press
to select the Input 1 Coupling menu.
•
Press
to highlight AC Coupling.
•
Press
to confirm;
•
Press
to select Input 2 READING.
•
Press
to select the Input 2 READING
•
Press
to highlight ACrms.
•
Press
to confirm;
•
Press
to highlight Input 2 Coupling.
•
Press
to select the Input 2 COUPLING.
•
Press
to highlight AC Coupling.
•
Press
to confirm;
•
Press
to return to SCOPE.
•
Press
to select RANGE 1 or RANGE 2.
•
Press
to select RANGE 1 or press
•
Press
to select the ranges mentioned in the table.
•
Press
when RANGE, RANGE 1, or RANGE 2 is not highlighted.
•
Press
to select
range shown under the appropriate column.
mark changes to I.
mark changes to I.
mark changes to I.
mark changes to I.
to select RANGE 2.