Input 1 And 2 Ac Voltage Accuracy Test - Fluke 43B Service Manual

Power quality analyzer
Hide thumbs Also See for 43B:
Table of Contents

Advertisement

Table 4-2. Volts DC Measurement Verification Points
Sensitivity
Input 1
Input 2
5 mV/div
5 A/div
10 mV/div
10 A/div
20 mV/div
20 A/div
50 mV/div
50 mV/div
100 mV/div
100 A/div
200 mV/div
200 A/div
500 mV/div
500 A/div
1 V/div
1 kA/div
2 V/div
2 kA/div
5 V/div
5 kA/div
10 V/div
10 kA/div
20 V/div
20 kA/div
50 V/div
50 kA/div
100 V/div
100 kA/div
1)
The 500V and 1250V range will be tested in Section 4.5.13
2)
Due to calibrator noise, occasionally OL (overload) can be shown.

4.5.9 Input 1 and 2 AC Voltage Accuracy Test

Dangerous voltages will be present on the calibration source
and connecting cables during the following steps. Ensure that
the calibrator is in standby mode before making any connection
between the calibrator and the test tool.
Proceed as follows to test the Input 1 and 2 AC Voltage accuracy:
1. Connect the test tool to the 5500A as for the previous test (see Figure 4-5).
5500A output,
V DC
[mV or V/div]
15 mV
14.4 to 15.6
30 mV
29.3 to 30.7
60 mV
150 mV
148.7 to 151.3
300 mV
298.0 to 302.0
500 mV
497.0 to 503.0
-500 mV
-497.0 to -503.0
0 mV
1.5V
1.487 to 1.513
3V
2.980 to 3.020
5V
4.970 to 5.030
-5V
-4.970 to -5.030
0V
-0.005 to +0.005
15V
14.87 to 15.13
30V
29.80 to 30.20
50V
49.70 to 50.30
-50V
-49.70 to -50.30
0V
-0.05 to + 0.05
150V
148.7 to 151.3
300V
298.0 to 302.0
Warning
4.5 Input 1 and Input 2 Tests in the SCOPE MODE.
DC Reading
Input 1
[A or kA/div]
2)
14.88 to 15.13
2)
29.80 to 30.20
59.2 to 60.8
59.65 to 60.35
148.7 to 151.3
298.0 to 302.0
497.0 to 503.0
-497.0 to -503.0
-0.5 to + 0.5
1.487 to 1.513
2.980 to 3.020
4.970 to 5.030
-4.970 to -5.030
-0.005 to +0.005
14.87 to 15.13
29.80 to 30.20
49.70 to 50.30
-49.70 to -50.30
-0.05 to +0.05
148.7 to 151.3
298.0 to 302.0
Performance Verification
Input 2
2)
2)
-0.5 to + 0.5