3. Set the 5500A to source a 1V, 1 MHz fast rising (rise time ≤ 1 ns) square wave
4. Set the 5500A to operate (OPR).
5. Press
6. Wait until the display shows
7. When you are finished, set the 5500A to Standby.
8. Continue at Section 5.6.3.

5.6.3 Gain DMM (Gain Volt)

Proceed as follows to do the Gain DMM calibration.
1. Press
2. Connect the test tool to the 5500A as shown in Figure 5-5.
5-10
Figure 5-4. 5500A Scope Output to Input 1
(SCOPE output, MODE edge).
to start the calibration.
The Delta T gain, Trigger Delay (CL0720), and Pulse Adjust Input 1 (CL0640) will
be calibrated.
Dangerous voltages will be present on the calibration source
and connection cables during the following steps. Ensure that
the calibrator is in standby mode before making any connection
between the calibrator and the test tool.
to select the first calibration step in Table 5-3.
Pulse Adj A (CL 0640):READY
Warning
ST8004.WMF
.