•
Press
•
Press
•
Press
•
Press
•
Press
•
Press
•
Press
•
Press
•
Press
•
Press
•
Press
4. Set the 5500A to source a sine wave, to the first test point in Table 4-5 (NORMAL
output, WAVE sine).
5. Observe the Input 1 and Input 2 main reading and check to see if it is within the
range shown under the appropriate column.
6. Continue through the test points of table 4-5.
7. When you are finished, set the 5500A to Standby.
Table 4-5. Volts Peak Measurement Verification Points
5500A output, Vrms (sine)
1.768 (5V peak)
4.5.12 Input 1 and 2 Phase Measurements Test
Proceed as follows:
1. Connect the test tool to the 5500A as for the previous test (see Figure 4-5).
2. Select the AUTO test tool setup:
•
Press
•
Press
•
Press
3. Select DC coupling for Input 1 and 2; Reading Phase for Input 2:
•
Press
•
Press
•
Press
•
Press
•
Press
to highlight DC Coupling.
mark changes to I.
to confirm;
select Input 2 READING.
to select the Input 2 READING.
to highlight Peak m/m.
mark changes to I.
to confirm;
to highlight Input 2 Coupling.
to select the Input 2 COUPLING.
to highlight DC Coupling.
mark changes to I.
to confirm;
to return to SCOPE.
5500A Frequency
1 kHz
to select the MENU.
till SCOPE is highlighted.
to select SCOPE mode
to select menu SCOPE SETUP.
to highlight Input 1 Coupling.
to select the Input 1 Coupling menu.
to highlight DC Coupling.
mark changes to I.
to confirm;
4.5 Input 1 and Input 2 Tests in the SCOPE MODE.
Reading 1
4.50 to 5.50
Performance Verification
Reading 2
4.50 to 5.50
4
4-21