POWER
SUPPLY
POSITIVE PULSE PERFORMANCE TEST
4-7.
With test equipment connected a s in Figure 4-2, proceed a s follows:
a.
Set Pulse Generator to output a positive-going 5-volt/lO ns pulse.
Set Pulse Generator repetition rate to approximately one-pulse-per-second. The probe
b.
indicator. should flash once every second.
c.
Disconnect test equipment.
PULSE
GENERATOR
DC
KEEP GROUND LEADS SHORT
Figure 4-7. Logic Level Performance Test Setup
SCOPE
1
Figure 4-2. Poistive Pulse Performance Test Setup
KEEP GROUND LEADS SHORT